![China Electronics Technology Instruments AV3672 Series Скачать руководство пользователя страница 104](http://html1.mh-extra.com/html/china-electronics-technology-instruments/av3672-series/av3672-series_user-manual_2601790104.webp)
4 Measurement Setting
4.6 Trigger Mode
100
Fig. 4.18 Trigger Setting Dialog Box
4.6.3 [Trigger] Dialog Box
1) Trigger setting options
a) Trigger source zone
The source of the channel trigger signal is determined by trigger source setting of the trigger source zone. The
effective trigger signal is generated when sweep is not done by the vector network analyzer. Three trigger sources
can be selected:
internal
,
manual
and
external
. The set trigger source will be used as the trigger source of all
channels.
➢
[Internal] (default)
1) Trigger signal generation is controlled automatically by the analyzer.
2) One trigger signal is generated immediately after one measurement.
➢
[Manual]
The manual trigger signal is generated in the following methods.
1) Click the
[Manual]
button in the
[Trigger]
dialog box.
2) Click the
[Manual
item in the
[Trigger]
sub-menu under the
[Stimulus]
menu of the menu bar.
➢
[External]
1) The trigger signal is input by
[External]
of the BNC connector on the rear panel.
2) The signal is TTL level.
3) Level trigger: the high or low level can be set in the
[External]
dialog box.
4) The pulse width is 1μs at and must not be longer than the sweep time (otherwise, the trigger may be enabled
several times).
b) Range zone
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...