background image

Chapter 2

 

Components Identification 

CMN034-002 

原本

(Original instructions)

 

39 

E

nglis

h

 

2.4

 

Display Indications 

2.4.1

 

About Modes 

 

 

2.4.2

 

While Stopped 

The Up/Down arrow buttons cycle through indications. 

 

 

 

 

Normal mode

 

ON button 

OFF button 

Page 44 

MODE SELECT button

 

Page 43 

ENTER button 

Held down for 3 sec 

Main power 

switch ON

 

Page 43 

OFF button 

Held down for 3 sec 

During 

stoppage

 

During 

operation

 

Ver*.**          ID.** 

Program version, RS485 communication ID 

Содержание CBA-750AT3-HC-DSA-V1

Страница 1: ...referred to as this manual and operate it after often reading this manual Please keep this manual carefully to be able to use it at any time English Dust Collector for Laser Marker MODELS CBA 750AT3...

Страница 2: ...ectively The dust collector for laser marker is an energy saving type clean box that realizes air technology in a compact body Please read this instruction manual thoroughly and handle this machine co...

Страница 3: ...7 2 2 1 CBA 750AT3 HC DSA V1 7 2 3 AT3 8 2 4 9 2 4 1 9 2 4 2 9 2 4 3 10 3 11 3 1 11 3 1 1 11 3 1 2 12 3 2 12 3 3 13 4 14 4 1 14 4 2 14 4 2 1 14 4 2 2 15 4 2 3 15 4 3 16 4 4 16 4 5 17 4 5 1 17 4 5 2 17...

Страница 4: ...2 6 24 6 1 24 6 1 1 24 6 1 2 26 6 1 3 27 6 2 27 6 2 1 RS485 27 6 2 2 27 6 3 27 7 28 7 1 28 7 2 28 7 3 29 7 3 1 CBA 750AT3 HC DSA V1 29...

Страница 5: ...1 CMN034 002 Translation of the original instructions 3 1 1 1 1 2 10 60 80 1 3 0 40 80 100 1 000...

Страница 6: ...1 4 1 4 2 OFF 1 5...

Страница 7: ...1 CMN034 002 Translation of the original instructions 5 1 6...

Страница 8: ...2 6 2 2 1 100V 200V T 220 230V CE 1 1 2 1 1 HEPA 1 2 3m 100V 1 200V T 1 220 230V CE 1 1 100 115V...

Страница 9: ...2 CMN034 002 Translation of the original instructions 7 2 2 2 2 1 CBA 750AT3 HC DSA V1 1 1 2 1 AT3 BOX HEPA...

Страница 10: ...2 8 2 3 AT3 EL OLED No 1 7 2 4 9 4 14 1 1 1 1 OFF 3 3 3 13 ON ENTER 3 3 3 13 4 14 MODE SELECT 4 14...

Страница 11: ...2 CMN034 002 Translation of the original instructions 9 2 4 2 4 1 2 4 2 ON OFF 14 13 ENTER 3 ON 13 OFF 3 Ver ID RS485...

Страница 12: ...2 10 2 4 3 OP kPa SP kPa DP kPa EP kPa Blower Runtime h 17 Total h Moter...

Страница 13: ...3 CMN034 002 Translation of the original instructions 11 3 3 1 3 1 1 0 40 80 RH REMOVE 1 1 1...

Страница 14: ...3 12 3 1 2 10 1 2 3 2 1 ON RS485 ID 2 AT3 ON 3 4 1 7 3 ON OFF 30 3 ON OFF 1 1 Ver ID...

Страница 15: ...nslation of the original instructions 13 3 3 WARN4 1 2 3 ENTER 3 Initial DP Get Y 4 ENTER MODE SELECT 5 1 7 Initial DP Check 6 Initial DP Entry WARN6 OFF 3 1 OFF 3 Initial DP Clr Y 2 ENTER MODE SELECT...

Страница 16: ...DE SELECT 4 2 4 2 1 14 4 2 2 15 4 2 3 15 4 2 1 1 2 Shaking Set 3 ENTER 4 Shaking Manual 5 ENTER 6 Shaking Start Y 7 ENTER 8 MODE SELECT Com Setting Volume Down ST Other Setting Error Data 16 16 17 17...

Страница 17: ...anslation of the original instructions 15 4 2 2 20 180 20 1 2 Shaking Set 3 ENTER 4 Shake Time sec 5 ENTER 6 7 ENTER 8 MODE SELECT 4 2 3 0 60 60 1 2 Shaking Set 3 ENTER 4 Interval min 5 ENTER 6 7 ENTE...

Страница 18: ...bps 19200bps 38400bps 57600bps 115200bps 9600bps 8bit 8bit 8bit 9bit 8bit 1 2 1 1 2 Com Setting 3 ENTER 4 5 ENTER 6 7 ENTER 8 MODE SELECT 4 4 WARN4 1 2 Volume Down ST 3 ENTER 50 3 Down to 50 4 30 70 5...

Страница 19: ...17 4 5 2 17 4 5 1 Runtime 1 2 Other Setting 3 ENTER Time Adjust 4 Runtime Reset 5 ENTER Reset Runtime Y ENTER 6 MODE SELECT 4 5 2 7 8 Other Setting 9 ENTER Time Adjust 10 Setting Reset 11 ENTER 1 YES...

Страница 20: ...5 18 5 5 1 WARN2 1 5 1 1 1 1 1 2 2 1 3 1 4 2 1...

Страница 21: ...5 CMN034 002 Translation of the original instructions 19 5 1 2 2 2 1 2 2 3 2 4 2 5 2 6 2 7 4 3 2 5 1 3 HEPA 1 4 2 HEPA HEPA 2...

Страница 22: ...5 20 3 4 5 HEPA 6 5 2 1 Littell fuse 218 250V 10A 1 2 5 3 1 1 1 1 5 4...

Страница 23: ...Translation of the original instructions 21 5 5 No 5 5 2 22 5 6 23 5 5 1 1 No No 2 MODE SELECT No 3 No 4 No 5 5 2 22 5 MODE SELECT 6 ENTER MODE SELECT ON WARN6 WARN7 ERR04 2 ERR04 No 1 WARN4 2 ERR04...

Страница 24: ...5 22 5 5 2 No ERR03 5 6 23 ERR04 5 6 23 ERR06 5 6 23 ERR07 F RAM F RAM ERR08 BCC WARN1 5 6 23 WARN2 5 6 23 WARN3 5 6 23 WARN4 5 6 23 WARN5 5 6 23 WARN6 13...

Страница 25: ...5 CMN034 002 Translation of the original instructions 23 5 6 EL ON ON 5 2 20 1 2 3 4 1 4 ON OFF 30 1 2 1 2 5 1 18 5 1 18 5 1 18 5 1 18...

Страница 26: ...6 24 6 6 1 6 1 1...

Страница 27: ...6 CMN034 002 Translation of the original instructions 25 1 5V 0 2V DC 1 2...

Страница 28: ...26 6 1 2 1 1 AT3 2 0 5V GND 1 5V 4 7k 3 LO NPN 3 LO NPN 3 LO NPN WARN2 WARN4 WARN5 ERR 1 2 0 5V 5 0V DCV 1 0 6 1 1 2 1 2 1 7 3 1 8 2 3 4 2 4 2 9 5 3 0 3 5 6 3 6 4 1 7 4 2 5 0 4 3 50V 100mA 1 2 33 0 1...

Страница 29: ...original instructions 27 6 1 3 ON OFF ON OFF 6 1 2 26 ON OFF 6 1 2 ON OFF 6 2 RS 485 RS EN ON OFF 6 2 1 RS485 4 3 4 3 16 6 2 2 6 3 mm 38 50 65 75 CBA 750AT3 HC DSA V1 FRJ D 38 35 108 FRJ D 50 35 108...

Страница 30: ...50 60Hz 1 3m3 min 9 5kPa 49 63dB 28 5kg CBA 750AT3 HC DSA V1 T 200V 5 0A 9 7kPa CBA 750AT3 HC DSA V1 CE 220 230V 4 7A 9 6kPa 1 1m A dB 2 3 100 115V 7 2 CBA 750AT3 HC V1 T CE 1 3 6 HDF 3030 120 F1 ZEO...

Страница 31: ...7 CMN034 002 Translation of the original instructions 29 7 3 7 3 1 CBA 750AT3 HC DSA V1...

Страница 32: ...30 12 7 2 7 2 28...

Страница 33: ...s in MODE SELECT Mode 44 4 2 Shaking Setting mode 44 4 2 1 Manual Setting Mode for Shaking 44 4 2 2 Time Setting Mode for Shaking 45 4 2 3 Interval time Setting Mode for Auto shaking 45 4 3 Communicat...

Страница 34: ...7 6 2 Communication Function 57 6 2 1 RS485 Communication 57 6 2 2 Ethernet 57 6 3 Removable Flange 57 Chapter 7 Appendix 58 7 1 Specifications 58 7 2 Consumables List 58 7 3 Electrical Diagram 59 7 3...

Страница 35: ...with flammable explosive or corrosive mist smoke or gases CAUTION This device is designed for installation in a cleanroom or a clean factory Avoid installation in other areas such as outdoors Ensure a...

Страница 36: ...to suck dry dust without potential dust explosion Be sure to connect the ground wire CAUTION Do not move the device while in operation Use the device in locations of pollution degree 2 Use a power sup...

Страница 37: ...MN034 002 Original instructions 35 English 1 6 Safety Label Locations High voltage warning label High temperature warning label Do not disassemble label Name plate label High voltage warning label Nam...

Страница 38: ...tects electrical parts from dust 1 Activated carbon cassette Adsorbs odor 1 Exhaust HEPA filter Cleans exhaust 1 Removable flange Connects a suction duct an exhaust duct 2 Power cable 3 m Connects to...

Страница 39: ...ust and fume Secondary filter Protects electrical parts from dust Shaking motor chamber Alleviates clogging in the primary filter AT3 panel operation panel Operates the device Main power switch Turns...

Страница 40: ...of the Left or Right arrow button changes the suction power to the next lower or higher level During MODE SELECT mode each pressing of the Left or Right arrow button moves the cursor left or right to...

Страница 41: ...odes 2 4 2 While Stopped The Up Down arrow buttons cycle through indications Normal mode ON button OFF button Page 44 MODE SELECT button Page 43 ENTER button Held down for 3 sec Main power switch ON P...

Страница 42: ...pressure Differential pressure Exhaust pressure Blower C Temperature around blower Runtime h Accumulated run time resettable Indicates run time accumulated since the last resetting Accumulated run ti...

Страница 43: ...ion Ambient temperature 0 to 40 C Ambient humidity 80 RH or lower without dew condensation Ambient conditions Indoors not exposed to direct sunlight free of corrosive flammable gases oil mist and dust...

Страница 44: ...is secured on the suction port 2 Connect the suction tube to be prepared separately to the removable flange IMPORTANT Ensure appropriate piping during operation The piping should be as short as pract...

Страница 45: ...alternates between the Initial DP Check message and the suction power level and differential pressure values 6 When the registration completes the display shows the Initial DP Entry message and the de...

Страница 46: ...Manual Setting Mode for Shaking Shaking is performed manually when the unit is not in operation 1 Move to the MODE SELECT mode 2 Press the Up Down arrow buttons to move to the shaking setting mode Sha...

Страница 47: ...lue 7 Press the ENTER button to determine the value 8 To exit this mode press the MODE SELECT button to return to normal mode 4 2 3 Interval time Setting Mode for Auto shaking The cumulative time unti...

Страница 48: ...number check screen appears showing the current settings 4 Press the Up Down arrow buttons to move to the item you want to set 5 Press the ENTER button The setting screen appears 6 Press the Up Down a...

Страница 49: ...lume down alert timing and resetting the accumulated run time 1 Move to the MODE SELECT mode 2 Press the Up Down arrow buttons to move to the other setting mode Other Setting 3 Press the ENTER button...

Страница 50: ...rs a WARN2 warning appears Replace the primary filter 5 1 1 Replacing the Primary Filter IMPORTANT The primary filter should be replaced in an area large enough to open the suction side filter case 1...

Страница 51: ...5 Remove the secondary filter 6 Install a new secondary filter 7 Install the filter presser plate removed in step 4 and the shaking guard removed in step 3 using screws close the suction side filter c...

Страница 52: ...crewdriver or other thin tipped tool to take it out carefully 2 Remove the fuses out of the fuse box and replace with new ones 5 3 Daily Inspection Inspection item Frequency Description Filter case Be...

Страница 53: ...ly Error No 2 Press the MODE SELECT button to move to the error history mode If there are two or more error warning occurrences the number of occurrences precedes the error number 3 If there are two o...

Страница 54: ...al Contact CHIKO AIRTEC WARN1 Internal temperature high Temperature around the blower is close to the error threshold Flashing Remains operational Follow the remedies for in 5 6 Troubleshooting Page 5...

Страница 55: ...e motor may be active Turn off the main power and after 30 minutes resume the operation Primary or secondary filter not in place Install the primary and secondary filters correctly Snap locks disengag...

Страница 56: ...Chapter 6 Useful Utilization Optional 54 Chapter 6 Useful Utilization Optional 6 1 Remote Cable 6 1 1 Standard Connection Diagram...

Страница 57: ...Dust side Customer side pin pin Measuring instruments etc Monitor Add a protection circuit so that the rating is not exceeded Dust side Customer side Recommended DC voltage current 1 2 of maximum rati...

Страница 58: ...occurs Green Operation output signal 3 Outputs LO during operation and HI when stopped Green white Abnormal output signal 3 Outputs HI during operation and LO when an error occurs Filter clogging out...

Страница 59: ...n power is turned off 6 2 Communication Function With the optional communication board set Model RS 485 or RS EN it is possible to retrieve the information like the ON OFF operation and filter cloggin...

Страница 60: ...is measured with a hose connected to the suction port at a distance of 1 m from the device on the A scale dB 2 2 7 2 9 m3 min When zeolite is adhered to the filter 3 Can be used in the area of 100 11...

Страница 61: ...Chapter 7 Appendix CMN034 002 Original instructions 59 English 7 3 Electrical Diagram 7 3 1 CBA 750AT3 DSA HC V1...

Страница 62: ...a CHIKO AIRTEC designated sales agent Blemish contamination or other appearance change that may occur during use Replacement of any consumable or accessory or use of any part not designated by CHIKO...

Страница 63: ...MEMO...

Страница 64: ...CHIKO AIRTEC CO LTD 562 0012 2 27 24 2 27 24 Hakushima Minoh Osaka 562 0012 Japan TEL 81 072 720 5151 FAX 81 072 720 5133 URL http chiko airtec jp...

Отзывы: