S423/C/OPT Optical oxygen probe – Data Sheet 

13 

 
 

ENTER 

Calibration of the sensor

 

 

This program step allows the calibration of the oxygen sensor. 
The calibration must necessarily be carried out:  

 Every time the sensor cap 9400410134 or 9400410135 is replaced 

 When starting after a long period of non-use 

 Whenever discrepancies occur, according to a known value 

For proper operation it is necessary, in addition to the cases mentioned 
above, to verify calibration or recalibrate the probe periodically. 
The frequency of this operation will be determined by the user, taking into 
account the type of application and the type of electrode used. 

 

NOTE 

To  perform  a  calibration  in  oxygen  saturated  solution  immerse  the 
probe in a container of clean water and wait 10 minutes for stabilization. 
Then continue with the Auto Calibration of the sensor 

 

 
 

 
 

 
 
 

 
Wait  for  the  values  (O

2

  and  °C)  to  stabilize,  then  press  ENTER;  if  the 

procedure is successful the display will show "Calibration OK”  
 
If ”Faulty Probe” is displayed, we recommend: 

 

To check the removable protection cap  

 

To assure the integrity of the membrane placed on the top of the 
probe  

 

To check the cable integrity, the correct connection to the instrument 
and on the probe. 

 

Automatic Calibration 
 
Press ENTER 
 
pO2 : 21.96% 
 

          Temp: 29.3°C 

Содержание S423/C/OPT

Страница 1: ...S423 C OPT SONDA OSSIGENO OTTICA OPTICAL OXYGEN PROBE MANUALE TECNICO TECHNICAL MANUAL P N Rev 0 Ver 1 0 EDIZIONE Febbraio 2014 EDITION February 2014...

Страница 2: ...ity to set salinity and barometric pressure for the compensation of the oxygen value Operating principle A relation exists between the oxygen concentration in the sample and the luminescence intensity...

Страница 3: ...e intensity in absence of oxygen Luminescence decay time in presence of oxygen 0 Luminescence decay time in absence of oxygen KSV Stern Volmer constant quantifies the quenching efficiency and therefor...

Страница 4: ...nstructions Electric installation Turn off the instrument connect the cables of the probe to the terminal block of the meter Please observe the colors shown on the sticker located under the electronic...

Страница 5: ...7connector for the connection to the 42 and 50 series units and can be provided with the appropriate probe holder S315 O available in various tube lengths Together with the probe holder is provided a...

Страница 6: ...the probe periodically The frequency of this operation will be determined by the user taking into account the type of application and the type of electrode used NOTE To perform a calibration in oxygen...

Страница 7: ...ss containing the polymer see figure below Furthermore it is necessary to pay particular attention to the substances with which the part of the probe deputed to measurement can come into contact for e...

Страница 8: ...tection IP68 Sensor cable Cable 10m integral Power Supply 12 24Vdc Signal interface RS 485 Modbus RTU Protocol Water move No necessary move Compensation of temperature Via internal NTC Luminophore dia...

Отзывы: