Guidance and Declaration - Electromagnetic Immunity
The Health Monitor is intended for use in the specified electromagnetic environment. The
customer or the user of the Health Monitor should assure that it is used in such an environment
as described below.
Immunity test
IEC60601
test level
Complianc
e level
Electromagnetic environment –
guidance
Conduced RF
IEC61000-4-6
3 Vrms 150
kHz to
80 MHz
outside ISM
bands
3 Vrms 150
kHz to
80 MHz
outside ISM
bands
Portable
and
mobile
RF
communications equipment should be
used no closer to any part of the
system, including cables, than the
recommended separation distance
calculated from the equation
appropriate for the frequency of the
transmitter.
Recommended
separation
distances:
P
d
2
.
1
Radiated RF
IEC61000-4-3
3 V/m 80
MHz to
2.5 GHz
3 V/m 80
MHz to
2.5 GHz
Recommended separation distances:
80
MHz
~
800
MHz:
P
d
2
.
1
800MHz-2.5GHz:
P
d
3
.
2