19
Harmonic
emissions
IEC61000-3-2
Class A
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings
used
for
domestic
purposes.
Voltage Fluctuations / Flicker
Emissions IEC 61000-3-3
Complies
Guidance and Declaration - Electromagnetic Immunity
The Health Monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the Health Monitor should assure that it is used in such an environment.
Immunity test
IEC60601
test
level
Compliance level
Electromagnetic
environment
-
guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be
wood, concrete
or ceramic tile. If
floors are
covered
with
synthetic material,
the
relative
humidity should
be at
least 30 %.
Electrical
fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
±
1
kV
for
input/output lines
Mains
power
quality should be
that of a typical
commercial
or
hospital
Surge
± 1 kV line(s) to
± 1 kV line(s) to