– 7 –
beneath the transducer, and that the transducer is seated flat against the material. If the
condition persists, it may be necessary to select a different transducer (size or frequency)
for the material being measured. See section 4.1 for information on transducer selection.
While the transducer is in contact with the material that is being measured, the TI-25PX
will perform four measurements every second, updating its display as it does so. When
the transducer is removed from the surface, the display will hold the last measurement
made.
2.9 Top End Cap
The top end cap is where all connections are made to the TI-25PX. The diagram above
shows the layout and description of the connectors.
Transducer Connectors
Refer to Diagram: The transducer connectors and battery cover/probe zero disk are
located on the TI-25PX’s top end cap. The transducer connectors are of type Lemo “00”.
NOTE:
There is no polarity associated with connecting the transducer to the TI-25PX, it
can be plugged into the gauge in either direction.
Probe Zero Disk & Battery Cover
Refer to Diagram: The Battery cover is the large round disk shown in the diagram.
NOTE:
This same disk is also used as a probe zero reference disk. Simply remove the
cover when replacing the batteries (2 AA cells). When performing a manual probe zero
function, simply place the transducer on disk making firm contact. Important: Be sure
the battery polarity is correct, which can be found on the back label of the TI-25PX.
NOTE:
Rechargeable batteries can be used, however they must be recharged outside of
the unit in a standalone battery charger.