I N T RODU CT I ON

The Electromatic Model TI-007 is a precision Ultrasonic Micrometer. Based on the 
same operating principles as sonar, the TI-007 is capable of measuring the thickness 
of various materials with accuracy as high as ± 0.0001 inches, or ± 0.001 millimeters. 
The principle advantage of ultrasonic measurement over traditional methods is that
ultrasonic measurements can be performed with access to only one side of the material
being measured.

This manual is presented in three sections. The first section covers operation of the 
TI-007, and explains the keypad controls and display. The second section provides
guidelines in selecting a transducer for a specific application. The last section provides
application notes and a table of sound velocity values for various materials.

Electromatic maintains a customer support resource in order to assist users with questions
or difficulties not covered in this manual. Customer support may be reached at any of the
following:

Electromatic Equipment Company
600 Oakland Avenue, 
Cedarhurst, NY 11516 – USA

Telephone: (516)295-4300
Facsimile: (516) 295-4399

Online: www.checkline.com
Email: [email protected]

N OT ES

Содержание TI-007

Страница 1: ...ion and Preparation of Surfaces 2 6 Selecting Measurement Mode 2 7 Calibration 3 0 TI 007 Modes 8 3 1 Scan Mode 3 2 Alarm Mode 3 3 Diff Mode 3 4 S 232 Port 4 0 Transducer Selection 10 5 0 Appendix A Product Specification 11 6 0 Appendix B Application Notes 12 7 0 Appendix C Sound Velocities of Common Materials 13 8 0 Warranty Information 14 OI UT003 ...

Страница 2: ...ed in three sections The first section covers operation of the TI 007 and explains the keypad controls and display The second section provides guidelines in selecting a transducer for a specific application The last section provides application notes and a table of sound velocity values for various materials Electromatic maintains a customer support resource in order to assist users with questions...

Страница 3: ... TI 007 is not in calibration mode the DOWN arrow key switches the display backlight between three available settings OFF will be dis played when the backlight is switched off AUTO will be displayed when the backlight is set to automatic mode and ON will be displayed when the backlight is set to stay on In the AUTO setting the backlight will illuminate while the TI 007 is making a measurement and ...

Страница 4: ...al storage device via the TI 007 s RS232 port Refer to page 9 for an explanation of how to use this feature The Display The numeric portion of the display consists of four complete digits preceded by a leading 1 and is used to display numeric values as well as occasional simple words to indicate the status of various settings When the TI 007 is displaying thickness measurements the display will ho...

Страница 5: ...cer is the business end of the TI 007 It transmits and receives ultrasonic sound waves that the TI 007 uses to calculate the thickness of thematerial being measured The transducer connects to the TI 007 via the attached cable and one coaxial connector The transducer must be used correctly in order for the TI 007 to produce accurate reliable measurements Below is a short description of the transduc...

Страница 6: ...l inhibit proper transducer coupling Extremely rough surfaces such as the pebble like finish of some cast iron will prove most difficult to measure These kinds of surfaces act on the sound beam like frosted glass on light the beam becomes diffused and scattered in all directions In addition to posing obstacles to measurement rough surfaces contribute to excessive transducer wear particularly where...

Страница 7: ...s characteristics It may be necessary to experiment with a variety of transducers in order to find one that works well for a given job ELECTROMATIC can provide assistance in choosing a transducer and offers a broad selection of transducers for evaluation in specialized applications 7 large ranges The TI 007 provides two simple methods for setting the sound velocity described in the following pages...

Страница 8: ... 4 Assuming ALAr value flashing IN is displayed use the UP and DOWN arrow keys to scroll to the desired Lo alarm value 5 Press the SEND key to select the Lo value entered The Hi value will now be displayed Repeat step 4 select the Hi alarm value 6 The TI 007 is now ready to perform measurements using the Alarm feature Differential Mode In the Quality Control environment it is sometimes necessary t...

Страница 9: ...C E Q U I P M E N T C O I N 600 Oakland Ave Cedarhurst NY 11516 U S A TEL 516 295 4300 FAX 516 295 4399 CHECK LINE INSTRUMENTS CHECK LINE BY ELECTROMATIC MODEL TI 007 ULTRASONIC M ICROM ETER OPERATING INSTRUCTIONS ...

Отзывы: