CHC
CTS-112R4 user manual
15
1)
Enter Star (
★
) key mode.
2)
Press [F2] to enter to tilt correction settings.
3)
When the tilt of instrument exceeds the correction range, it needs leveling manually. Follow the
steps described in “Instrument Setup” to center the black dot, as shown on the right.
Single Axis: only corrects the vertical angle. Dual Axis: corrects vertical and horizontal angles.
4)
Press [F4] (↓) to display the tilt value on X (horizontal) and Y (vertical) direction. It needs leveling
manually. Rotate the screws on the tribrach to level the instrument. Press [ESC], return to Star key
mode. Press [F3] to disable the correction.
5)
If the tilt sensor is disabled, press [F1] (X-ON) or [F2] (XYON) to activate the correction function.