10
DATE:
09/
01
/00
TIME:
00:01:15AM
DATE:
09/12/15
TIME:
09:37:25AM
Press
to go to the next field, then
or
to define the
field. Press
again to go to the third field, then
or
to
define it. Press
to validate.
Proceed in the same way for the time.
CHANNEL 1:
ENABLED
INPUT:
±100 mV
UNITS:
CHANNEL 1:
TRIGGER:
RISING
SCALE LOW:
0.0 mV
= 0.0 mV
CHANNEL 1:
SCALE HIGH:
100.0 mV
= 100.0 mV
CHANNEL 1:
EQUIVALENCE:
1.00 Pulse
= 1.00 Wh
CHANNEL 1:
5. CONFIGURING THE MEASUREMENT CHANNELS
Press
twice. From this screen you can activate or deactivate
measurement channel 1 and choose the type of input:
pulse
event (discrete)
4-20mA current
±10V, ±1V, or ±100mV voltage
Press
to go to the next screen and enter the unit of channel
1 in 4 characters.
If you do not specify a unit, the default values will be:
mV for ±100 mV
V for ±10V and ±1 V
mA for 4-20 mA
Wh for pulse
For an event input, you can choose whether triggering is on
the positive-going or negative-going front.
Press
. The next 2 screens are used to define transformation coefficients for the current and voltage
inputs. Only one screen for a pulse input.