5.2.4
MEASUREMENT
■
Place the conducting wire of the circuit to be measured in the clamp and close
the clamp. If the clamp is incorrectly closed, the
icon is displayed.
■
If necessary, use the
HOLD
key to freeze the measurement. See §4.6.
■
If necessary, use the
MEM
key to store the measurement. See §4.8.2.
5.2.5
MEASUREMENT RESULT
First screen
Once the measurement has stabilized,
the display unit displays the 1st screen,
which indicates:
■
The leakage current.
■
The loop impedance referred to the
chosen frequency.
The impedance is measured only if the
leakage current is less than 10A. In the
10A-40A range, only the current is
displayed; the
NOISE
symbol blinks
and the impedance is replaced by
dashes.
Fig. 14
Second screen
Press
► to display the 2nd screen,
which indicates the contact voltage
(product ZxI).
Fig. 15