44
5.6.
REINITIALISING THE INSTRUMENT
■
■
■
■
■
To reinitialise all the test parameters, please follow the procedure below:
1. Switch off the instrument by setting the ON/OFF key to the OFF position.
2. Press the SET UP key and keep it pressed while switching on the instrument.
Machine Tester followed by Hard Reset are displayed for a few moments, then the main menu opens.
NOTES :
The
reinitialised values of the adjustable parameters correspond to the requirements of standard EN 60204.
■
■
■
■
■
List of the test parameters per function, with their initial value.
Function
Parameter
Range of adjustment
Initial value
or possible values
U
N
Test voltage
1, 1.25, 1.5 kV AC
1 kV
1, 3, 5, 10, 20, 30, 40, 50, 60, 70,
HV
Imax
Tripping current
80, 90, 100, 150, 200, 250, 300,
500 mA
333, 350, 400, 450, 500 mA
t
Timer
1s...1min...5 min (in steps of 1 s and 10 s)
1 s
∆
U
∆
U
max
Max permitted
5.0 V (0.50 mm
2
), 5.0 V (0.75 mm
2
),
voltage drop
3.3 V (1.0 mm
2
), 2.6 V (1.5 mm
2
),
3.3 V (1.0 mm
2
)
1.9 V (2.5 mm
2
), 1.4 V (4.0 mm
2
),
1.0 V
≥
6.0 mm
2
t
Timer
1 s...15 s
10 s
R
10A
AC
R
max
Max permitted
10 m
Ω
...1000 m
Ω
(in steps of 10 m
Ω
)
*m
Ω
resistance
or
*
m
Ω
(no limit)
t
Timer
1s...15 s
10 s
U
N
Test voltage
500, 1000 V DC
500 V DC
R
min
min. allowed
0.2 M
Ω
...1 M
Ω
(in steps of 0.1 M
Ω
)
M
Ω
insulation
1 M
Ω
...10 M
Ω
(in steps of 1 M
Ω
)
1 M
Ω
resistance
10 M
Ω
...500 M
Ω
(in steps of 10 M
Ω
)
or
*
Ω
(no limit)
t
Timer
2 s...1min...10 min (in steps of 1 s and 10 s)
15 s
∆
t
Syst
Measuring
2-point (1 s), 4-point (5 s)
2-point (1 s)
system
t
Timer
ON ou OFF
ON
ON/OFF (M/A)
Toutes
Baud
Baud rate
300, 600, 1200, 2400, 4800
4800
fonctions
Buzzer
ON/OFF
ON
Contrast
0%...100% (in steps of 2%)
50%
■
■
■
■
■
To reinitialise all the test parameters and erase all stored data, please follow the procedure below :
1. Switch off the instrument by setting the ON/OFF key to the OFF position.
2. Press the SET UP and MEM/MR keys and keep them pressed while switching on the instrument.
Machine Tester then Hard Reset followed by Clear Memory are displayed for a few moments, then the main menu opens.
NOTES :
The
reinitialised values of the adjustable parameters correspond to the previous table, in accordance with
standard EN 60204.