![Chauvin Arnoux C.A 5275 Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/chauvin-arnoux/c-a-5275/c-a-5275_user-manual_2594519022.webp)
22
22
4.2.5 Resistance
Special reference conditions: the input (+, COM) must not have been overloaded following the accidental application of
voltage on the input terminals when the switch is on the
Ω
or T ° position. If this is the case, the return to normal can take
about ten minutes.
Range
Specified
measurement range
Resolution
Uncertainty (±)
Measurement
current
Open circuit
voltage
600
Ω
0 – 600.0
Ω
0.1
Ω
1 % R + 3 D
≈
1 mA
< 5 V
6 k
Ω
0 – 6.000 k
Ω
0.001
k
Ω
1 % R + 2 D
≈
120
μ
A
60 k
Ω
0 – 60.00 k
Ω
0.01
k
Ω
≈
12
μ
A
600 k
Ω
0 – 600.0 k
Ω
0.1
k
Ω
≈
1.2
μ
A
6 M
Ω
0 – 6.000 M
Ω
0.001
M
Ω
1.5 % R + 3 D
≈
120 nA
60 M
Ω
0 – 60.00 M
Ω
0.01 M
Ω
3 % R + 5 D
≈
30 nA
4.2.6 Audible continuity
Response time < 100 ms
Range
Resolution
Uncertainty (±)
Voltage in open circuit
Meas. current
600
Ω
0.1
Ω
Beep triggered
< 30
Ω
+ 3
Ω
< 5 V
≈
1,1 mA
4.2.7 Diode test
Range
Resolution
Uncertainty (±)
Voltage in open circuit
Meas.
current
6 V
1 mV
Beep triggered
< 40 mV + 10 mV
< 5 V
< 1.1 mA
4.2.8 Operation of the beep
Beep indicating a valid key
Æ
high sound
4 kHz, 100 ms
Beep indicating an invalid key
Æ
low sound
1 kHz, 100 ms
Successive beeps for 30 seconds ending with a long beep indicating that the
instrument is being shut down
Æ
medium sound
2 kHz 100 ms
3 successive beeps with a space of 1 second in between (beep beep beep - gap -
beep beep beep) indicating that the danger-level threshold has been exceeded
Æ
medium sound
2 kHz, 100 ms
2 successive beeps (beep beep) indicating MIN, MAX recording:
Æ
medium sound
2 kHz, 100 ms
Current > 10 A
4 kHz, 100 ms