AmpFlex
®
Current Probe
15
3.3 TYPICAL RESPONSE CURVES
0.00 %
-0.50 %
-1.00 %
-1.50 %
-2.00 %
-2.50 %
-3.00 %
1 Hz
10 Hz
100 Hz
1000 Hz
10 kHz
100 kHz
Typical Accuracy % vs. Frequency @ 100 A, 1 mV/A Output
Frequency
0.00 %
-1.00 %
-2.00 %
-3.00 %
-4.00 %
-5.00 %
-6.00 %
-7.00 %
1 Hz
10 Hz
100 Hz
1000 Hz
10 kHz
100 kHz
Typical Accuracy % vs. Frequency @ 100 A, 10 mV/A
Frequency
%
± 10
± 9
± 8
± 7
± 6
± 5
± 4
± 3
± 2
± 1
± 0
0.1 A
0.5 A
1 A
5 A
10 A
100 A
20 A ... 3000 A
10 kA ... 30 kA
Typical Accuracy % per Range to 100 A
Amps