53
Intelect
®
Advanced Laser Module
TROUBLESHOOTING
ERROR MESSAGES
CODE
NUMBER
TYPE
MESSAGE
PROBABLE CAUSES
POSSIBLE REMEDIES
130
Warning
Another treatment is running
while attempting to set up
and perform a Laser Therapy
treatment.
A. Allow existing treatment to complete before starting Laser Therapy.
B. If no other treatment is running, reset Therapy System by turning main
power switch Off and On.
131
Warning
Treatment Room Door Lockout is
breached.
A. Make certain Treatment Room Door is completely closed.
B. Make certain the Lockout cable is connected to the system.
C. Replace Lockout to System cable with a known good cable.
D. Contact department responsible for installation of the Treatment
Room Door Lockout mechanism for maintenance or repair.
E. Send Laser Module to Factory for Service.
132
Warning
Attempted to start a laser
treatment but no laser applicator
is plugged in.
A. Connect desired Laser Applicator to the system.
B. If Applicator is connected, reset Therapy System by turning main
power switch Off and On.
C. Connect a known good Laser Applicator.
D. Send Laser Module to Factory for Service.