37
Intelect
®
Advanced Laser Module
11.
Press the START button. A message displays to alert you to the fact that the
laser is armed. The unit will beep two times to count down the number of
seconds left until the applicator is ready. Press the Pause/Resume button to
begin the treatment.
After you press the Pause/Resume button, an orange light will illuminate on
the back of the laser applicator to indicate that laser output is being
distributed.
NOTE:
When treatment time has expired, a tone will sound.
The therapy can be interrupted at any time by pressing the Pause/Resume
button on the back of the applicator, the STOP button (on the Operator
Interface), or the PAUSE button (on the Operator Interface).
When the STOP button is pressed, the applicator stops emitting laser energy,
and the Completed Treatment Review window displays. Press the button next
to Start New Treatment to begin a new therapy session (refer to
step 2 on
page 31
), or press Save to Patient Card to transmit the parameters of the
treatment to the patient card. For more information on the Save to Patient
Card option, see the unit’s user manual.
OPERATION
STARTING, STOPPING, AND INTERRUPTING THERAPY (continued)