Intelect® Transport Combo
55
OPERATION
STARTING, STOPPING, AND INTERRUPTING COMBINATION THERAPY CONTINUED
21.
If you want to use Channel 2 for additional electrotherapy, continue with step 22.
If you do not want to use additional electrotherapy, skip to step 25.
22.
Press the Up and Down Arrow buttons to highlight Stim Channel 2.
23.
Press
the
Enter
button.
The
Select
Waveform
screen
displays.
24.
Follow steps 8-12 of the section entitled
"Starting, Stopping, and Interrupting
Electrotherapy"
on
pages 43-47
.
25.
You may make parameter changes and stop or pause therapy on Channel 1, Channel
2, or Ultrasound by highlighting the appropriate form of therapy, and making the
necessary
changes.
26.
When you have completed treatment, remove all electrodes from the patient.
27.
Turn system power “OFF” by pressing the Power On/Off button.
The unit beeps once and the blue light on the Power On/Off button flashes
intermittently.