SPECIFICATIONS
25
Premodulated current is a medium frequency waveform. Current
comes out of one channel (two electrodes). The current intensity
is modulated: it increases and decreases at a regular frequency
(the Beat Frequency).
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-100 mA
Carrier Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2500 Hz
Beat Fixed (Sweep Off ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 Hz
Sweep Low Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-199 Hz
Sweep High Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-200 Hz
Cycle Time . . . . . .Continuous, 5/5, 4/12, 10/10, 10/20, 10/30, and 10/50
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 min
*CC= Constant Current
CV= Constant Voltage
WAVEFORM SPECIFICATIONS
Premodulated
Interferential current is a medium frequency waveform. Current is
distributed through two channels (four electrodes). The currents
cross each other in the body at the area requiring treatment.
The two currents interfere with each other at this crossing point,
resulting in a modulation of the intensity (the current intensity
increases and decreases at the beat frequency).
Carrier Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2000-5000 Hz
Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 Hz
Sweep Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 Hz
Sweep Low Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-199 Hz
Sweep High Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-200 Hz
Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-100 mA
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 min
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Interferential