*Accuracy: Resistance measurements within ± 10% (E11 and 

greater ± 20%), complies with ESD-S4.1. Open circuit voltages 

of 10 volts ± 5% and 100 volts ± 5% exceeds requirements of 

ESD-S4.1
Meter Weight: 439 grams
Dimensions: 203 mm L x 109 mm W x 41 mm H
Test Switch: Press the red test button for 15-20 second 

electrification period.  Meter will immediately display surface 

resistance exponent via LED.  LCD will display temperature 

(both Fahrenheit & Celsius), then humidity, and after taking 

numerous readings and making numerous calculations display 

surface resistance mantissa. 

Note: Reporting test results, you may want to record the 

resistance, humidity, temperature, and test voltage.

Maintenance

Your Surface Resistance Test Kit will require little maintenance, 

and there are no user serviceable parts. If your Meter requires 

service beyond cleaning the

Electrodes or replacing the batteries, please contact the factory.

This product utilizes a high frequency switching circuit to step 

up the 3 volts from the batteries to the 100 volt test level.  

Some users are able to discern a slight hum or buzzing.  This is 

perfectly normal and should not be considered a flaw or defect.

Calibration*

The Surface Resistance Test Kit is calibrated to NIST traceable 

standards.  Most users require calibration annually.

Please call our Customer Service Department at 00 44 (0) 

1892-665313 for details. In-house calibration can be performed 

by using 1% resistors in each of the meter ranges. Simply 

attach the resistors to the enclosed cords using grounding clips 

and recording the meter display. Keep the cords separated.   

Should adjustment be necessary, it is recommended that the 

unit be returned to the factory as access to internal adjustments 

requires special equipment and test cables.

EQUIPMENT:

•  Digital Multimeter - accurate to 1.25% @ 10VDC and 

100VDC

•  Fixed Decade Box value 10E3 - 10E12 - accurate to ±2.5%, 

except at 10E11 and 10E12 (±5%)

•  Thermometer - accurate to ±1°F

•  Humidity meter - accurate to 2%

•  Test leads

•  99% Isopropyl alcohol and cleaning wipes

SET UP:

A .Test Area

 - Area needs to be free of any high voltage 

transformer or power supply. Not under any type of fluorescent 

lighting or high power lighting.

B. Worksurface

 - needs to be covered with conductive mat at 

1.0 x 10E3 or less, connected to earth ground.

C. Technician

 - needs to be grounded with zero ohms resistor 

to earth ground.

D. Decade Box

 - needs to be grounded to earth ground.

*Accuracy is measured after normalizing the instrument at 

70-80 °F and 30-50% RH for a minimum of 4 hours.

NORMALIZATION OF SURFACE RESISTANCE TESTER

Temperature inside testing area needs to be 75°F @ ±6.6% 

40% to 60% RH. Testers need to stay at constant temperature 

75°F @ ±6.6% for about 2 hours for proper reading. Testers 

cannot be inside objects, enclosed boxes, containers or cases 

unit is supplied with (temperature inside case will differ from 

outside temperature, cases will act like an insulator to the 

Testers), Testers will have to be stationary in testing area for 

about 2 hours with no dramatic temperature changes.

TESTING OF SURFACE RESISTANCE TESTER

•  When testing use only the leads supplied with the tester. DO 

NOT TEST WITH RESISITIVITY ATTACHMENT, Resitivity 

Attachment will give a much higher reading and is only 

designed for spot checking.

•  With 10V/AUTO switch down, press TEST button - voltage 

between the two leads should be 10V ±5%

•  With 10V/AUTO switch up, press TEST button - voltage 

between the two leads should be 100V ±5%

•  Using the cleaning wipes and 99% isopropyl, clean around 

banana jack and mono jack where leads connects to, oil from 

human fingers can alter accuracy.

•  Make sure 10V/AUTO switch is set to AUTO (switch up). 

Testing each decade starting from 1.0 x 10E12 and down, 

never start from 1.0 x 10E3.

Temperature 

Fahrenheit 

= 75°F ±6.6% 

 

Celsius   

= 23.8°C ±10.2%
Relativity Humidity ±10 Digits 

A. 1.0 x 10E12

+ 20%  LED = 12 Yellow   

Mantissa 1.20

0% 

LED = 12 Yellow   

Mantissa 1.00

- 20%   LED = 11 Yellow   

Mantissa 8.00

B. 1.0 x 10E11

+20%  LED = 11 Yellow   

Mantissa 1.20

0% 

LED = 11 Yellow   

Mantissa 1.00

-20% 

LED = 10 Green   

Mantissa 8.00

C. 1.0 x 10E10

+10%  LED = 10 Green   

Mantissa 1.10

0% 

LED = 10 Green   

Mantissa 1.00

-10% 

LED =   9 Green   

Mantissa 9.00

D. 1.0 x 10E9

+10%  LED =   9 Green   

Mantissa 1.10

0% 

LED =   9 Green   

Mantissa 1.00

-10% 

LED =   8 Blue   

Mantissa 9.00

E. 1.0 x 10E8

+10%  LED =   8 Blue   

Mantissa 1.10

0% 

LED =   8 Blue   

Mantissa 1.00

-10% 

LED =   7 Blue   

Mantissa 9.00

Содержание 99105

Страница 1: ...ance Measurement Test method A 1 Floor working surface or storage rack working surface A 2 Seating A 3 Garments A 5 4 Gloves finger cots and tools Neither 20 mm plates nor clip connectors provided Versatility Measures Rp Rg or Resistivity requires optional Resistivity Attachment item 99107 Quick Checks Surface resistance exponent number illuminates immediately i e 8 10E8 ohms or 100 000 000 ohms M...

Страница 2: ...ponent or power of the number is displayed immediately with LED illuminating number If 8 is illuminated the measurement is in the 10E8 or 10E8 ohm range To make a measurement the button is to be depressed for 15 20 seconds for the unit to make numerous readings and calculations Testing in accordance with EN 61340 5 1 requires 15 to 20 seconds of electrification in contrast most analog type meters ...

Страница 3: ...ectrode on different sections Clean surface for Laboratory test but do not initially clean surface for installed products if fails clean and retest Laboratory Test Procedure Guideline For laboratory test of ESD Working surfaces Floor Materials Footwear Garments or Seating best advice is to follow test methods of Annex A of EN 63140 5 1 The test specimens and electrodes should be cleaned twice with...

Страница 4: ...cally perform a lot sampling at incoming inspection Floor Per EN 61340 5 1 paragraph 10 1 These audits shall be performed at intervals not exceeding those defined by the ESD co ordinator Users typically perform these quarterly Seating Per EN 61340 5 1 paragraph 10 1 These audits shall be performed at intervals not exceeding those defined by the ESD co ordinator Users typically perform these quarte...

Страница 5: ...ate to 2 5 except at 10E11 and 10E12 5 Thermometer accurate to 1 F Humidity meter accurate to 2 Test leads 99 Isopropyl alcohol and cleaning wipes SET UP A Test Area Area needs to be free of any high voltage transformer or power supply Not under any type of fluorescent lighting or high power lighting B Worksurface needs to be covered with conductive mat at 1 0 x 10E3 or less connected to earth gro...

Страница 6: ...d or replaced at Charleswater s option free of charge Call Customer Service at 00 44 0 1892 665313 for a Return Material Authorisation RMA and for proper shipping instructions and address You should include a copy of your original packing slip invoice or other proof of purchase date Any unit under warranty should be shipped prepaid to the Charleswater factory Warranty replacements will take approx...