background image

Unit 17. Millbrook Business Park, Sybron Way • Crowborough, East Sussex TN6 3JZ United Kingdom

Phone: 00 44 (0) 1892-665313 • Internet: 

Charleswater.co.uk

PPE-5105.E

 Page 3 of 4

© 2013 Charleswater

5.  Re-assemble the tester.

Specifications

Output Voltage: 2.5 volts DC @ 10mA
Dimensions: 12 cm x 4.5 cm x 2 cm
Operating Temperature: 0° - 40°C

Calibration

The Low Resistance Tester is calibrated to standards 

traceable to NIST.  Frequency of recalibratrion should be 

based on the critical nature of those ESD sensitive items 

handled and the risk of failure for the ESD protective 

equipment and materials.  In general, we recommend that 

calibration be performed annually.
Calibration is performed by placing various resistors between 

the probe tip and the lead wire.
1.  Attach the alligator clip to the lead wire.
2.  Set the Low Resistance Tester to the 1 ohm test setting.
3.  Attach the lead of a 1 ohm resistor to the alligator clip.
4.  Firmly place the tester’s probe tip on the opposite resistor 

 lead.
5.  Press and hold the test button.  The PASS LED should 

 illuminate.
6.  Repeat the procedure with a 1.2 ohm resistor.  The PASS 

  LED should not illuminate.
7.  Repeat this process with the ranges listed in the table 

  below.  If the unit fails calibration, check and replace the 

 battery.

Test Range Setting

PASS

FAIL

1 ohm

0 to 1 ohm

≥ 1.2 ohms

2 ohms

0 to 2 ohms

≥ 2.2 ohms

20 ohms

0 to 20 ohms

≥ 22 ohms

Maintenance

BATTERY REPLACEMENT

1.  Properly ground yourself using a wrist strap.
2.  Remove the 2 screws located at the back of the tester.

Figure 5.  Locating the enclosure screws

Figure 6.  Removing the back cover

3.  Remove and turn over the circuit board.
4.  Locate and replace the battery (3 Volt; Model CR2032).

First, place 

battery under 

prongs

Next, press here to snap 

battery in place

Figure 7.  Replacing the battery

Содержание 99036

Страница 1: ...ering and desoldering equipment the measurement techniques can be used for other electrical hand tools The working part of AC powered tools should be capable of providing a conductive path to ground New powered hand tools such as soldering irons typically should have a tip to ground resistance of less than 1 0 ohm NOTE This resistance may increase with use but should be less than 20 0 ohms for ver...

Страница 2: ...and the PASS LED does not illuminate Example of Test Range Uses 1 OHM Range ESD Technical element Grounding Conductors Banana Jacks Common Ground Points Earth Bonding Points Grounding Blocks Mat Ground Cords ANSI ESD S6 1 Grounding sections 6 4 Technical Elements 6 4 1 6 4 2 and 6 4 3 New AC Powered Hand Tools ESD Handbook ESD TR20 20 section 5 5 2 2 Electrical Hand Tools 2 OHM Soldering Irons ESD...

Страница 3: ...ire 2 Set the Low Resistance Tester to the 1 ohm test setting 3 Attach the lead of a 1 ohm resistor to the alligator clip 4 Firmly place the tester s probe tip on the opposite resistor lead 5 Press and hold the test button The PASS LED should illuminate 6 Repeat the procedure with a 1 2 ohm resistor The PASS LED should not illuminate 7 Repeat this process with the ranges listed in the table below ...

Страница 4: ... proof of purchase date Any unit under warranty should be shipped prepaid to the Charleswater factory Warranty replacements will take approximately two weeks Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES EXPRESSED AND IMPLIED INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED The express warranty wil...

Отзывы: