SECTION 1
–
INSTALLATION 1-1
Section 1
–
Installation
Unpacking the Instrument
Remove the instrument from the packing crate carefully. The unit comes fully equipped with
all the necessary components and ordered spare parts. Make sure that no parts are lost when
discarding the packing materials. Place the instrument on a firm table, close to the water
source and required electrical outlets.
After the instrument is removed from the shipping crate, the equipment and spare parts
should be checked against the packing list to ensure that all parts have been received and
none are damaged.
Note: File an insurance claim with your freight carrier if damage has occurred
during shipping. Verify all parts shown on the enclosed packing list have
been received. If items are missing, please notify Chandler Engineering
immediately.
Utilities Required
Your unit will require dry, oil-free compressed shop air (not instrument quality) of 100 to
130 psi / 689
–
896 kPa, and a water supply of 20 to 80 psi / 138
–
552 kPa. The unit is
capable of operating in ambient temperatures from 50°F to 120°F / 10°C - 49°C.
Tools/Equipment Required
An English unit mechanics tool set is adequate for the installation, operation, and
maintenance of the instrument. No special tools are required.
This unit is supplied with an accessory kit, which includes the necessary hardware for the
water, air, and electrical hook-ups. The water and air hose may be cut to length and the
appropriate barbed fittings inserted into the hose and clamped into place.
Caution: The laboratory electrical power wiring must be capable of a 45-ampere
load and comply with local electrical codes. The instrument must be
securely connected to an appropriate earth ground. The ground wire must
have a larger diameter than that of the supply voltage conductors.
Setting up the Instrument
The air, drain, and water connections are made at the lower rear of the instrument. The inlets
are 1/4in. female pipe thread. The installation kit, provided with the instrument, contains the
necessary hose and fittings to make the connections. Electrical connections are made using
the 50-ampere rated connector and power cord provided with the instrument. Be certain that
all power wiring is rated for the 45-ampere load. The wiring must be installed in accordance
with local electrical codes, and the entire instrument must be securely connected to a separate
ground.
Содержание AMETEK 5617
Страница 3: ......
Страница 9: ...P 4 PREFACE This page is intentionally left blank ...
Страница 23: ...5 2 SECTION 5 REPLACEMENT PARTS This page is intentionally left blank ...
Страница 29: ...5 6 SECTION 5 REPLACEMENT PARTS This page is intentionally left blank ...
Страница 31: ...6 2 SECTION 6 DRAWINGS AND SCHEMATICS This page is intentionally left blank ...
Страница 32: ......
Страница 33: ...CHANDLER ENGINEERING ...
Страница 36: ...TB1 TB2 ...
Страница 37: ...CHANDLER ENGINEERING ...
Страница 38: ......
Страница 41: ...cut out postcards on dotted lines ...