Chapter 5
SAFETY FEATURES
[87]
Open End of the Motor Housing
5.
Holding the motor assembly firmly in one hand with the open end of the motor adapter
housing facing down, sharply rap the flat surface of the motor adapter housing on a thin
piece of cardboard placed on a solid surface. See Figure 11-11. Only one or two sharp raps
should be needed to dislodge the wave generator bearing, the flexspline, the motor shaft
key, and the Lexan
®
washer. Any clean, slightly cushioning surface will work for this step,
but wood is not recommended because small wood chips could get lodged in the ball
bearings.
Dislodging the Wave Generator Bearing
NOTE:
The flexspline is the slightly flexible spline that fits on the outside of the wave
generator bearing. It has external teeth.
6.
Verify that all four components: the wave generator bearing, the flexspline, the motor shaft
key, and the Lexan
®
washer, dislodge from the motor adapter housing. Set them aside.
Flexspline
Snap
Ring
Motor
Adaptor
Housing
Fixed
Spline
Wave
Generator
Bearing
Rap Sharply
One or Two
Times
Содержание Ametek Quizix 5000 Series
Страница 8: ...Chandler Engineering 5000 SERIES PUMP MANUAL 6 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 14: ...Chandler Engineering 5000 SERIES PUMP MANUAL 12 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 18: ...Chandler Engineering 5000 SERIES PUMP MANUAL 16 Stand Mounted Q5200 System...
Страница 19: ...Chapter 2 SYSTEM SETUP 17 Expanded Front View of a Stand Mounted Q5200 System...
Страница 20: ...Chandler Engineering 5000 SERIES PUMP MANUAL 18 Stand Mounted Q5400 System...
Страница 34: ...Chandler Engineering 5000 SERIES PUMP MANUAL 32 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 40: ...Chandler Engineering 5000 SERIES PUMP MANUAL 38 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 48: ...Chandler Engineering 5000 SERIES PUMP MANUAL 46 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 66: ...Chandler Engineering 5000 SERIES PUMP MANUAL 64 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 70: ...Chandler Engineering 5000 SERIES PUMP MANUAL 68 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 120: ...Chandler Engineering 5000 SERIES PUMP MANUAL 118 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 126: ...Chandler Engineering 5000 SERIES PUMP MANUAL 124 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 127: ...125 Appendix A A Declaration of Conformity is found on the following page...
Страница 152: ...Chandler Engineering 5000 SERIES PUMP MANUAL 150 THIS PAGE INTENTIONALLY LEFT BLANK...