![Ceyear 1465 Series Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/ceyear/1465-series/1465-series_programming-manual_2584913059.webp)
3 SCPI
3.3 Instrument subsystem command
52
Setting format:
[:SOURce]:LIST:FILL:EXECute
Parameter:
Example:
:LIST:FILL:EXECute Accomplish automatic filling of list.
Key Entry:
【
Sweep
】
—>[List Sweep]—>[Auto Fill]
Note:
For setting only.
[:SOURce]:LIST:DELete <Mode>
Function:
This command is used for deleting the points in the list of List Sweep.
Setting format:
[:SOURce]:LIST:FILL:EXECute
Parameter:
Discrete data. Options for deleting the list points are:
CURRent Current point
ALL All points
Example:
:LIST:DELete ALL Delete all points in the list.
Key Entry:
【
Sweep
】
—>[List Sweep]—>[Edit List]—>[Del All]
Note:
For setting only.
3.3.5 LFOutput subsystem
The following commands are used for setting operation modes:
[:SOURce]:LFOutput:AMPLitude ........................................................................ 52
[:SOURce]:LFOutput:FREQuency ...................................................................... 52
[:SOURce]:LFOutput:FREQuency:ALTernate .................................................... 53
[:SOURce]:LFOutput:FREQuency:ALTernate:AMPLitude:PERCent ................. 54
[:SOURce]:LFOutput:NOISe ............................................................................... 54
[:SOURce]:LFOutput:RAMP ................................................................................ 54
[:SOURce]:LFOutput:SHAPe .............................................................................. 55
[:SOURce]:LFOutput:STATe ............................................................................... 55
[:SOURce]:LFOutput:SWEep:TIME .................................................................... 55
[:SOURce]:LFOutput:AMPLitude <Ampl>
Function:
This command is used for setting the output signal amplitude from the LF output BNC
(Bayonet Nut Connector) of the signal generator
Setting format:
[:SOURce]:LFOutput:AMPLitude <val>(unit:VPP|Mvpp|VRMS)
Query format:
[:SOURce]:LFOutput:AMPLitude?
Parameter:
<Ampl> LF output signal amplitude.
Range: 2.000Vpp[0.002Vpp,4.000Vpp].
Example:
:LFOutput:AMPLitude 1Vpp LF output signal amplitude is set to 1 Vpp.
Reset status:
2.000Vpp
Key Entry:
【
Frequency
】
—>[LF Out]—>[LF Ampl]
[:SOURce]:LFOutput:FREQuency <Frequency>
Function:
This command is used for setting LF output. It should be noted that, when LF Waveform
Содержание 1465 Series
Страница 1: ...1465 series Signal Generator Programming Manual...
Страница 2: ......
Страница 5: ......
Страница 39: ......
Страница 127: ......