![cetoni NEMESYS XL Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/cetoni/nemesys-xl/nemesys-xl_hardware-manual_2584861011.webp)
NE
MESYS XL Manual
11
2.4.3.
Reasonably Foreseeable Misuse
The use of the device beyond its intended use may lead to dangerous or harmful
situations and must be avoided.
ATTENTION
The device must not be used as a medical product nor in medical
applications.
2.4.4.
Safety Measures
To ensure the safety of the user and a failure-free operation of the device, only
original parts as well as original accessories must be used. No damages can be
claimed when third-party components or accessories are used.
The device has been developed and designed to avoid harmful situations during
intended use. However, the following safety measures have to be observed to
exclude potentially remaining risks:
cetoni GmbH advices of the operator's responsibility during use. Relevant
laws and regulations at the location of use must be observed. In the
interest of safety, operators and users are responsible to adhere to all
relevant regulations.
The device must not be used as a medical product nor in medical
applications.
The device is designed and approved to work in fluidic systems, which fall
within the scope of Article 4 Paragraph 3 of the Pressure Equipment
Directive 2014/68/EU.
This means that the system may not exceed a maximum volume of 1 liter.
With the use of fluids from Group 1 according to Article 13 of the
Pressure Equipment Directive 2014/68/EU, the maximum allowable
system pressure is 200 bar. For fluids from Group 2 it is 1000 bar. If
different, product-specific values for the maximum pressure are given in
the section "Technical Data", these values must be complied with.