VC431
Instruction manual
29
2.6.3 Preliminary settings: Orientation of the accelerometer and
projection Time
In whole body (W) measurements the orientation of the accelerometer should be selected
(backrest position or seated position) for the correct adjustment of the reference axes.
Before starting to measure, check that the accelerometer orientation indicated by the
VC431
is correct (see 2.6.1).
To evaluate exposure to vibrations for measuring times below the exposure times the
VC431
calculates the projected daily exposure to vibration that corresponds to the daily exposure to
vibration that would exist if measurement time were the same as the projection time.
For this purpose, simply enter the projection time (tp) that corresponds to the forecast
exposure to noise time. By default, the
VC431
assigns 8 hours to the projection time, but this
value can be modified.
To modify orientation time and the tp value:
Make sure that the measurement has been stopped
Press button
Select the right orientation (backrest
or seat
) by means of
and
Validate with
The
VC431
will then highlight the value of the hours of the tp (projection time) in bold
Choose the numerical value by means of
and
Validate with
Repeat the process to modify the minutes.
The projection time tp (hh:mm) may take values:
o
From 0 to 24 for HH (Hours)
o
From 0 to 55 for mm (minutes) in 5-minute steps