Cerlic pHX Скачать руководство пользователя страница 16

                                                                                                           

 

- 16 - 

 

If smart calibration is used and BB2 recognizes the buffer, BB2 will suggest a pH 

value showing the text ”Accept smart calibration pH X.X”. You can choose to 

accept the value or to change it manually. 

 

If manual calibration is used BB2 will always ask for the value of the buffer, enter 

the value and press ENTER. 

 

The second buffer is not accepted if it is less than 1 pH from buffer 1, if so the 

second calibration will be repeated. 

 

When the calibration is finished the slope is shown in the display. "Slope XX%". 

This gives an indication of the shape of the electrode. A good electrode will have a 

slope between 93 and 102 %  

 

Press ENTER to finish the calibration.  

Adjusting ORP 

The ORP sensor is not calibrated, an offset adjustment is done instead. The offset 

adjustment can be done in  two ways automatic or manual. Irrespective of the way you 

chose, you need a buffer with a known potential. 
Remove the sensor and, if necessary the electrode as well, but make sure that the 

electrode is still connected to the signal amplifier. Wipe clean the electrode before 

adjustment. 

 

Spray the electrode using clean water, then immerse it in the first buffer, and wait 

for a stable reading.  

 

Enter the sensor menu. The output is frozen when the menu is opened. 

 

Open the “Calibrate” menu, then select "Adjust", chose "Set” and press ENTER. 

 

A window pops up asking for the ORP potential. Enter it and then press ENTER. 

BB2 will calculate the difference and use it as offset. 

14.

 

Scaling 

The “Scale / Alarm” menu allows the user to set the high and low boundaries for a 4-

20mA output signal.  In addition, this menu allows the user to set high and low alarms 

values that can be used to notify when solids have reached critical points. 

Max 

sets the 20 mA point output 

Min 

sets the 4 mA point output  (may be a negative value for special 

applications) 

High-Alarm 

sets the High Alarm set point  

Low-Alarm 

sets the Low Alarm set point 

Содержание pHX

Страница 1: ...2019 02 20 C60C5EN19 pHX ReX pH Redox Sensor...

Страница 2: ...e for cleaning solenoid valves 8 10 Maintenance 9 Error messages 9 Broken Electrode 9 Incorrect Measurements 9 11 Sensor display 10 12 Menu for Sensors 11 Menu for pHX 11 Menu for ReX 13 13 Calibratio...

Страница 3: ...de which in combination provide a voltage in linear proportion to the pH value Electrodes internally combined with a reference and measuring cell are divided into two categories gel electrodes and ref...

Страница 4: ...05264 Flushing system 1 meter excl solenoid valve P N 11205612 Flushing system 2 meter excl solenoid valve P N 11205903 Solenoid valve 220V for flushing P N 11705516A Incoming hose 16mm 5 8 to sensor...

Страница 5: ...ations where the level varies greatly a mounting kit with a float ball should be used Take care to mount the mounting kit in such a manner that it is easily accessible in conjunction with calibration...

Страница 6: ...s may need to be rinsed away with water If the electrode has holes for refilling the electrolyte inside the electrode should be checked and refilled The level should be at least 1 cm 1 2 below the ref...

Страница 7: ...dia Then tighten the nut with your hand If you have an electrode which does not have a hard wired cable the protective cover on top of the electrode should be removed and the enclosed plug in cable sh...

Страница 8: ...being the master If needed specify the extra freeze time in seconds If the sensor shall be cleaned as master specify the relay to be used according to the wiring inside the control box For example if...

Страница 9: ...calibration the pHX does not perceive a difference of buffers Make sure that you use two different buffers during calibration The electrode may be broken The electronic components in the sensor may b...

Страница 10: ...has two information pages the first one shows in addition to the measured value Max Min and average values for the last 24 hours The values are based on hour values which means they may jump when a ne...

Страница 11: ...he first buffer Sample 2 pH Actual pH for the second buffer Calibrated Date of last air calibration New eletcr Date when electrode was last changed Push ENTER to edit the date then ENTER again to stor...

Страница 12: ...eaning System Type Type of sensor read only Serial Serial number of the sensor read only SoftW Software version of the sensor read only Info Press ENTER to go to info read only menu This menu is for C...

Страница 13: ...eaning cycles Length sec 0 999 seconds duration of flushing cycle Freeze sec 0 999 seconds extra freeze time of output signal after a flushing cycle Relay 1 2 Along 1 or Along 2 Select relay to operat...

Страница 14: ...are version of the sensor read only Info Press ENTER to go to info read only menu This menu is for Cerlic internal use RV 0 mV Raw value for 0 mV RV 1000 mV Raw value for 1000 mV Ch1 Raw value ORP mV...

Страница 15: ...4 00 7 00 or 10 00 of a known temperature it is best to pour the buffers into plastic cups Enter the sensor menu The output is frozen when the menu is opened Open the Calibrate menu Select Calibrate t...

Страница 16: ...se you need a buffer with a known potential Remove the sensor and if necessary the electrode as well but make sure that the electrode is still connected to the signal amplifier Wipe clean the electrod...

Страница 17: ...ciple Electrode 12mm Cable connection 5 pin M12 plug Cable length 10 m 33 ft Cable material Polyurethane Enclosure IP65 ReX P N 11305598 Tube and electrode holder SIS2 343316SS Top PVC Dimensions 28 x...

Страница 18: ...3 68 EEC 89 336 EEC EMC Directive as amended by 92 31 EEC and 93 68 EEC The following standards and or technical specifications have been applied EN 61000 6 4 2001 Electromagnetic compatibility EMC Pa...

Страница 19: ...19 16 Dimensions...

Страница 20: ...20 Appendix 1 Optional slide rail holder...

Страница 21: ...roxide ions the output from the electrode is 0 0 mV and pH is 7 The millivolt signal is measured by pHX and the corresponding pH is calculated in BB2 mV pH H ions Mol l OH ions Mol l Example 414 0 1 0...

Страница 22: ...3 1 70 2 76 3 82 4 88 5 94 7 00 7 94 8 88 9 82 10 76 11 70 12 64 55 131 1 55 2 64 3 73 4 82 5 91 7 00 7 91 8 82 9 73 10 64 11 55 12 46 65 149 1 40 2 52 3 64 4 76 5 88 7 00 7 88 8 76 9 64 10 52 11 40 1...

Страница 23: ...n the main display Go to the sensor menu it is accessed by pressing ENTER for five seconds Select System and press ENTER Type _____________________________________________________ Serial _____________...

Страница 24: ...aning sub menu of the sensor menu the following parameters can be set Cleaner _____________________________________________________ Cleaning interval __________________________________________________...

Отзывы: