- 12 -
Scale / Alarm
Max
Measured value corresponding to 100%, equal to 20 mA output
signal (or 4 mA if 20-4 mA is used).
Min
Measured value corresponding to 0%, equal to 4 mA output
signal (or 20 mA if 20-4 mA is used).
Hi-Alarm
Level that activates an alarm when exceeded, 0 = not in use.
Low-Alarm
Level that activates an alarm when underpassed, 0 = not in use.
Alarm Relay
”-” ”#1”, ”#2”, or ”#1 and #2”. Check that the relay is not being
used for cleaning
System
Type
Type of sensor, read only
Serial
Serial number of the sensor, read only
SoftW
Software version of the sensor, read only
Info
Press “ENTER” to go to “info” read only menu. This menu is for
Cerlic internal use.
MS 1 mV
Millivolt value for the first calibration point
MS 2 mV
Millivolt value for the second calibration point
pH 1
pH for the first calibration point
pH 2
pH for the second calibration point
RV 0 mV
Raw value for 0 mV
RV 200 mV
Raw value for 200 mV
Ch1
Raw value
MS mV
Electrode output in millivolt
pH
Current measure value
Samp/s
Number of samples per second
Service
Not accessible for users.