- 7 -
O2X
DUO
EElk
•
If a new electrode is not to be installed, then the yellow plastic cap shall be
mounted to protect the electrode connection plug and the flushing nozzles on the
sensor.
Installing a new electrode/cell:
•
Loosen and remove the SS flushing head.
•
Remove the cable connector from the electrode/cell and clean the connector.
•
Use a new o-ring, 31.6x2.4 mm EPDM, on the electrode when exchanging the
electrode.
•
Plug the electrode/cell cable into the connector on the top of the electrode.
•
Install the new electrode/cell into the SS flushing head.
Don’t touch the cap
lens on the electrode/cell when installing.
•
For the optical cell, the grooves in the electrode must be adapted to the holes for
the flush pipes.
•
Tighten the SS flushing head and mount the protection plate, if needed. See
below the instructions on mounting the protection plate.
•
A new air-calibration must be performed whenever the electrode/cell is changed.
For the optical cell also a 0-calibration shall be done.
•
The date when the electrode/cell is changed shall be entered in the calibration
menu.
Mounting the protection plate
First: remove the small o-rings mounted in the groove at the top of the three flushing
nozzles. Bend the plate slightly and press it down between the flushing nozzles and into
the grooves. Confirm that the plate fits into the notches on the flushing nozzles. Do not
remove a flushing nozzle in order to mount the protection plate. The nozzles are
fastened with a special locking glue.
Wrench
Electrode
Flushing head