Tube placement Cytology & Cell wash rotors
Cytology Rotors
Manual Cell W ashing Rotor
ALWAYS
load your holders opposite to each other
ALWAYS
weigh samples to within
0.5
Gram. For best separations
ALWAYS
fit the lid supplied
Check
the two top Gaskets and replace regularly (Part Number 8055)
CHECK the rotor screw fixing tightness regularly
ALWAYS
load your tubes opposite to each other
ALWAYS
weigh samples to within
0.2
Gram. For best separations
Regularly smear a small amount of Silicon grease over the O rings
Check
the inner and outer O rings and replace regularly (Part Number 8066)
CHECK the rotor screw fixing tightness regularly
✔
✔
✘
✘
✘
✘
................................................................................................................................................................................
Always fit the lid supplied, as shown below
9