IMPORTANT
Read before using
Location
The Centrifuge should only be operated indoors.
The set up location must fulfil the following requirements
•
Check the Centrifuge and the Packaging for any shipping damage.
Inform the shipping company and Centurion Scientific immediately
•
Check the order for completeness, if not contact Centurion Scientific
•
A safety zone of at least 30Cm must be maintained around the Centrifuge
•
People and hazardous substances must be kept out of this zone
during centrifugation
•
The bench or supporting structure must be stable and free from resonance
•
The Bench or supporting structure must be suitable for horizontal use of the
centrifuge
•
The Centrifuge should Not be exposed to heat or strong sunlight
•
Ultra violet rays can damage the stability of plastics and coatings
•
Do not subject the Polypropylene, Delrin rotors or accessories to direct
sunlight
•
The Zone location must be well ventilated at all times
Transporting the Centrifuge
•
Always remove the Rotor before transportation
•
Due to its weight the Centrifuge should be carried by several people
•
Transport the Centrifuge upright and if possible in its packaging
•
Always lift the centrifuge at both sides (see below). NOT front to rear.
•
Store the original packaging, Keep this for future service to base , if required.