DRSplus
–
Operating Manual
54/58
Table 3
–
Electromagnetic Immunity (ISO 60601-1-2:2007 5.2.2.2)
Immunity Test
IEC60601 test
level
Compliance
Level
Electromagnetic environment guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
3 Vrms
150KHz to
80MHz
3V/m
80MHz to
2.7GHz
3Vrms
3V/m
Portable and mobile RF equipment should be used
no closer to any part of
drs
plus
, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d =
1.17√P
d =
1.17√P 80MHz to 800MHz
d =
1.17√P
800MHz to
2.5GHz
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic sight surveya,
should be less than the compliance level in each
frequency rangeb
Interference may occur in the vicinity of equipment
marked with the following symbol.
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflections from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcasts and TV broadcasts cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which
drs
plus
is used exceeds the applicable RF compliance level above,
drs
plus
should
be observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as re-orientating or relocating
drs
plus
.
b Over the frequency range 150Khz to 80MHz, field strengths should be less than 3V/m.
18.3
Immunity tests performance criteria
Function
Immunity tests performance criteria
Device operation -
main unit
During application of the test stimulus, any temporary cessation or
interruption of the intended operation remains within acceptable limits.