33 34
4.000uF
1nF
+3.0% of rdg + 5 dgts
40.00uF
10nF
100.0uF
0.1uF
+5.0% of rdg + 5 dgts
Input Protection:
250V dc or 250V ac rms.
Frequency
(Auto-ranging)
Range
Resolution
Accuracy
5Hz
0.001Hz
+1.5% of rdg + 5 dgts
50Hz
0.01Hz
500Hz
0.1Hz
+1.2% of rdg + 3 dgts
5kHz
1Hz
30.00kHz
10Hz
Sensitivity:
>5V RMS MIN.
Overload protection:
250V dc or ac rms.
Duty Cycle
Range
Resolution
Accuracy
0.1%~99.9%
0.1%
+1.2% of rdg + 4 dgts
Pulse width:
>100us, <100ms; (Continued next page)
Sensitivity:
<0.5V RMS
Overload protection:
250V dc or ac rms.
Period
Range
Resolution
Accuracy
2.0~ 20.0ms
0.1ms
+3% of rdg + 10 dgts
Overload protection:
250V dc or ac rms.
Pulse Width
Range
Resolution
Accuracy
2.0~ 10.0ms
0.1ms
+3% of rdg +10 dgts
Overload protection:
250V dc or ac rms
Temperature
Range
Resolution
Accuracy
-20oC~+760oC
1 oC
+3% of rdg +5dgts(Meter only,
probe accuracy not included)
-4 oF~+1400 oF
1oF
Sensor:
Type K Thermocouple
Diode Test
Test current
Resolution
Accuracy
0.3mA typical
1 mV
+10% of rdg + 5 dgts
Open circuit voltage:
1.5V dc typical
Overload protection:
250V dc or ac rms.
Audible continuity
Audible threshold: Less than 150
Ω
Test current: <0.3mA
Overload protection: 250V dc or ac rms.