FREQUENCY (DUTY CYCLE) MEASUREMENTS (ELECTRONIC)
1. Set the rotary function switch to the green
Hz/%
position.
2. Insert the black lead banana plug into the
negative
COM
jack and the red test lead
banana
plug into the positive
Hz
jack.
3. Touch the test probe tips to the circuit under
test.
4. Read the frequency on the display.
5. Press the
MODE
button to indicate ―
%
‖.
6. Read the % duty cycle in the display.
% 4
– 20mA MEASUREMENTS
1. Set up and connect as described for DC mA measurements.
2. Set the rotary function switch to the
4-20mA%
position.
3. The meter will display loop current as a % with 0mA=-25%,
4mA=0%, 20mA=100%, and 24mA=125%.
INSULATION RESISTANCE MEASUREMENTS
a)
Set the rotary function switch to the
INSULATION
position ,
and Press the
RANGE
button to chose one of the voltage which
display on the top left corner.
b)
Connect two testing lines to the tested
;
c)
Push down and hold the
―TEST‖ button /or press the ― LOCK‖
keystoke first and then the
―TEST‖ button, if the tested is
electriferous and its voltage ( AC/DC) is over 30V, it will refuse work
and no high-voltage testing occurs, simultaneity, it shows
―>30V‖ on
the LCD, the symbol
― ‖ flashes, and the buzzer warns frequently.
if the tested is diselectriferous or its voltage is lower than 30V, it will
enter into the formal testing process and brings the high-voltage. on
the primary display, the insulation resistance in M
Ω is indicated in-