5
Auto Power Off
To conserve battery life, the meter will automatically turn off after
approx. 30 minutes of non-use. When this happens, the state of the
meter is saved. In order to disable auto power off function, power on
the meter when any of the push function , except for HOLD, is
pressed down.The “APO” sign on the LCD panel indicates whether
the auto power-off function is enabled or not.
Operating Temperature:
0ºC to 40ºC (32ºF to 104ºF)
and Humidity below 80% RH
Storage Temperature:
-10ºC to 60ºC (14ºF to 140ºF)
and Humidity below 70% RH
Power source:
6x1.5V Size “AA” battery or Equivalent (DC9V)
Dimensions:
200(L) x 92(W) x 50(H) mm
Weight:
Approx 700g include battery
Accessories:
4 sets Test kits, 4pcs iron rods, 6pcs battery, Carrying
case, manual.
IV. Electrical Specifications
Accuracies are specified in the way:
± (…% of r…digits) at 23ºC ± 5ºC,below 80% RH.
Low resistance
Range
Resolution
Accuracy
Current
400mΩ
0.1mΩ
± (1%+10d)
200mA
4Ω
1mΩ
± (1%+5d)
20mA
40Ω
0.01Ω
± (1%+5d)
2mA