background image

 

POWER BUTTON 

I. 

Press the power button, power is on and the meter can 

measure. 

II. 

Press it again, Power is off. 

 

 

 

 

BATTERY REPLACEME 

I. 

As  battery  power  is  not  sufficient,  LCD  will  display 
“          ”  replacement  with  one  battery  type  9V  is 
required. 

II. 

Open  battery  cover,  then  take  out  the  battery  from 

instrument  and  replace  with  a  new  9-Volt  battery  and 

place the battery cover back. 

 

 
7. Common Sources of CO 

Common sources of potentially dangerous levels of CO are: 

·

  Poorly maintained furnaces, gas heaters, or fireplaces. 

·

  Dirty or plugged chimneys, or flue exhausts.5 

·

  Poorly maintained gas, oil, or kerosene appliances. 

·

  Internal  combustion  engines  (e.g.,  automobiles, 

lawnmowers, blowers). 

 

 

Содержание CO-181

Страница 1: ...1 Carbon Monoxide Meter Instruction Sheet...

Страница 2: ...ze the effects of CO poisoning 0 1PPM Normal background levels 9 PPM ASHRAE Standard 62 1989 for living areas 50 PPM OSHA enclosed space 8 hour average level 100 PPM OSHA exposure limit 200 PPM Mild h...

Страница 3: ...discontinuously with the Concentration of CO 4 Specifications Temperature Operating Storage 0 to 50 30 to 60 Operating humidity 0 99 Relative humidity non condensing Measurement range 0 to 1000PPM Mea...

Страница 4: ...4 5 Instrument Familiarization CO sensor LCD Display MODE button Alarm light Power button SEL button Battery door...

Страница 5: ...CD displays CO ppm and data Keep pressing SEL button over four seconds you can turn on off the backlight Press MODE button to enter maximum data mode Maximum data mode In this mode LCD displays the ma...

Страница 6: ...the position of storage data by pressing SEL button Recall alarm mode Press MODE button again to enter this mode the LCD displays the alarm data 30ppm When the measuring data is more than 30ppm and le...

Страница 7: ...7 middle of LCD will firstly with flicker then turn into zero and enter into the measure mode Measure mode Maximum data mode Recall data mode Store data mode Recall alarm mode Measuring Time mode...

Страница 8: ...off function in this mode While in other modes the meter will enable auto power off after about 15 minutes without pressing any button Set alarm In this mode you can turn on off the buzzer by pressin...

Страница 9: ...ed II Open battery cover then take out the battery from instrument and replace with a new 9 Volt battery and place the battery cover back 7 Common Sources of CO Common sources of potentially dangerous...

Страница 10: ...ems Gas furnaces Room heaters Oil natural gas or LPG liquefied petroleum gas 1 Cracked heat exchanger 2 Not enough air to burn fuel properly 3 Defective blocked flue 4 Maladjusted burner 5 Building no...

Страница 11: ...LPG 1 Not enough air to burn fuel properly 2 Defective blocked flue 3 Maladjusted burner 4 Building not properly pressurized Ranges Ovens Natural gas or LPG 1 Not enough air to burn fuel 2 Maladjuste...

Страница 12: ...12 exchanger 5 Cracked firebox...

Отзывы: