Slimline Slide Drying Hotplate
Symptoms
Possible Cause
Action Required
5.
Temperature of the hot plate
on the controller, is different to
the temperature of the surface
of the hotplate measured by a
reference probe.
A: Check correct probe is being
used for measurement and that
the probe is calibrated.
B: This is completely normal
and is due to position of the
heating element under the
hotplate.
C: Measure temperature at the
position where the instrument
is calibrated, using a calibrated
probe. If the temperature is
reading is significantly different,
the instrument may need to
be re-calibrated. Follow the
calibration instructions.
A: Check correct probe is being
used for measurement and that
the probe is calibrated.
B: Measure temperature at the
position where the instrument
is calibrated, using a calibrated
probe. If the temperature is
reading is significantly different,
the instrument may need to
be re-calibrated. Follow the
calibration instructions.
6.
Temperature of the hot plate
continues to rise when not
expected.
A: Instrument is still performing
tuning exercise after start up.
B: Actual hotplate temperature
is lower than the set
temperature
C: Temperature control circuit
fault.
A: Wait for tuning exercise to
finish, this takes roughly 10-15
minutes after start up.
B: Check the set temperature.
C: Instrument should be
checked by a competent
person.
Содержание JAY-0100-00A
Страница 1: ...Slimline Slide Drying Hotplate User Manual www CellPath com T 44 0 1686 611333...
Страница 16: ...www cellpath com T 44 0 1686 611333 16 User Manual Notes...
Страница 17: ...17 www cellpath com T 44 0 1686 611333 17 Slimline Slide Drying Hotplate Notes...
Страница 18: ...www cellpath com T 44 0 1686 611333 18 User Manual Notes...