Slimline Slide Drying Hotplate

Symptoms

Possible Cause

Action Required

5.

 Temperature of the hot plate 

on the controller, is different to 

the temperature of the surface 

of the hotplate measured by a 

reference probe.   

A: Check correct probe is being 

used for measurement and that 

the probe is calibrated. 
B: This is completely normal 

and is due to position of the 

heating element under the 

hotplate.
C: Measure temperature at the 

position where the instrument 

is calibrated, using a calibrated 

probe.  If the temperature is 

reading is significantly different, 

the instrument may need to 

be re-calibrated. Follow the 

calibration instructions.

A: Check correct probe is being 

used for measurement and that 

the probe is calibrated. 
B: Measure temperature at the 

position where the instrument 

is calibrated, using a calibrated 

probe.  If the temperature is 

reading is significantly different, 

the instrument may need to 

be re-calibrated. Follow the 

calibration instructions.

6.

 Temperature of the hot plate 

continues to rise when not 

expected.

A: Instrument is still performing 

tuning exercise after start up.  
B: Actual hotplate temperature 

is lower than the set 

temperature 
C: Temperature control circuit 

fault.

A: Wait for tuning exercise to 

finish, this takes roughly 10-15  

minutes after start up.  
B: Check the set temperature.
C: Instrument should be 

checked by a competent 

person.

Содержание JAY-0100-00A

Страница 1: ...Slimline Slide Drying Hotplate User Manual www CellPath com T 44 0 1686 611333...

Страница 2: ...86 611333 2 User Manual Contents 3 General Notes 3 Specimen Safety 4 User Safety 4 Power Lead and Connection to Electrical Supply 5 Specification 5 Location 6 Operating Instructions 7 Cleaning Instruc...

Страница 3: ...should be replaced with an equivalent power supply unit Before using any cleaning or decontamination method please refer to the Maintenance and Cleaning section to ensure the proposed method will not...

Страница 4: ...h will prolong the life of the instrument to ensure continued safety Do not touch any electrical contacts or open any closure plates RISK OF ELECTRIC SHOCK DO NOT Allow molten wax to accumulate on the...

Страница 5: ...limline Slide Drying Hotplate features a digital temperature control over temperature cut out and miniature circuit breakers for safety Dimensions Width 183mm Depth 450mm Height 85 5mm Capacity 28 Sli...

Страница 6: ...ssed for more than one second the value increases decreases rapidly and after two seconds pressed the speed increases even more to allow the desired values to be reached rapidly D Exiting the Set mode...

Страница 7: ...l quantity of mild domestic detergent applied with a soft synthetic sponge The microscope slide support rods can be removed to facilitate easier cleaning SCOURING PADS OR DE SCALING AGENTS MUST NOT BE...

Страница 8: ...ck temperature with a calibrated digital thermometer fitted with a surface probe The temperature should be measured in the centre of the hotplate back to front and side to side as shown in the image C...

Страница 9: ...splay will now show a number this number is the current adjustment offset applied to the temperature display This was the adjustment offset which was used when the slide drying hotplate was manufactur...

Страница 10: ...Thermocouple PTFE Insulated 1 Off Routine Inspection Recommendations CellPath Ltd recommend that a simple annual inspection be made for all CellTec laboratory equipment in order that any malfunction...

Страница 11: ...1333 11 Slimline Slide Drying Hotplate Wiring Diagram SW1 MCB x2 E N L BODY HOB R38 ELEMENT 1 ELEMENT 2 NEON 1 3 SSR 5 6 4 2 110V N AC AC DC DC N N 230V CUT OUT 120 C 1 2 3 4 5 6 7 8 11 12 OUT 2 OUT 1...

Страница 12: ...e 2 Power is supplied to the instrument but the hot plate does not heat Temperature does not rise on the controller and the orange heater light does not operate A Thermal cut out has triggered B Tempe...

Страница 13: ...instrument may need to be re calibrated Follow the calibration instructions A Check correct probe is being used for measurement and that the probe is calibrated B Measure temperature at the position w...

Страница 14: ...to or inscribed upon the goods have been removed covered up or defaced in any way Any goods or parts thereof which may require repair or replacement shall be repaired or replaced at the election of Ce...

Страница 15: ...ection and repair of the equipment will be issued prior to any work being carried out On return of the repaired equipment to the customer it is the customer s responsibility to ensure that the loan eq...

Страница 16: ...www cellpath com T 44 0 1686 611333 16 User Manual Notes...

Страница 17: ...17 www cellpath com T 44 0 1686 611333 17 Slimline Slide Drying Hotplate Notes...

Страница 18: ...www cellpath com T 44 0 1686 611333 18 User Manual Notes...

Страница 19: ...w Voltage Directive 2014 35 EU EMC Directive 2014 30 EU We confirm the declaration Nickel Electro Ltd Oldmixon Crescent Weston Super Mare North Somerset BS24 9BL United Kingdom Conforms with the requi...

Страница 20: ...CellPath Ltd 80 Mochdre Enterprise Park Newtown Powys SY16 4LE UK T 44 0 1686 611333 E sales CellPath com CellPath com 00 659 372 V3...

Отзывы: