05 Cleaning
1. Remove the printhead from the printhead mount. To prevent motor damage, use one hand to secure the
top of the printhead mount while sliding the printhead upward with the other hand.
2. Remove the nozzle with the wrench while the printhead is still warm.
3. Return the printhead to the mount and reheat the Thermoplastic Printhead.
4. Wait 15 minutes to ensure the granulate is fully melted.
5. Apply pressure to the printhead to extrude the excess material until only air is expelled (Figure 10).
6. Remove the printhead from the printhead mount.
7. Unscrew the top cap.
8. Gently insert the plunger retrieval tool into the cartridge and find the threaded hole in the plunger.
9. Turn the plunger retrieval tool clockwise at least five turns to screw into the plunger.
10. Firmly pull the plunger up to retrieve the plunger.
CAUTION:
Plunger and melted thermoplastic residue may be very hot. Touching hot surfaces and objects
can result in serious injury.
Содержание BIO X
Страница 1: ...BIO X Thermoplastic Printhead User Manual...
Страница 2: ......
Страница 4: ......
Страница 5: ...1 Package contents 01...
Страница 7: ...3 Technical specifications 02...
Страница 11: ...7 Safety information 03...
Страница 14: ...10...
Страница 15: ...11 Getting started 04...
Страница 25: ...21 04 Getting started Figure 8 Loading the Thermoplastic Print head with granulate A B C...
Страница 37: ...33 Cleaning 05...
Страница 40: ...36...
Страница 41: ...37 Printing optimization 06...
Страница 44: ...40...
Страница 45: ...41 Relevant G code commands 07...
Страница 48: ...44...
Страница 49: ...45 Frequently asked questions 08...
Страница 54: ...50...
Страница 55: ...51 Maintenance 09...
Страница 57: ...53 Appendix A Consumables...
Страница 60: ...www cellink com...