1. GENERAL INFORMATION

Thank you for your purchase of 

CEL-MAR Company

 product. This product has been completely tested and is covered by a two year

warranty on parts and operation from date of sale. 

If any questions or problems arise during installation or use of this product, please do not hesitate to contact Technical Support at +48

41 362-12-46 or e-mail [email protected].

1.1. WARRANTED INFORMATION 

ADA-401WP 

module is covered by a two year warranty from date of sale. The warranty does not cover damage caused from improper

use, materials consumption or any unauthorized changes. If the product does not function (is damaged), or not operate in accordance

with the instructions will be repaired.
All warranty and no warranty repairs must be returned with paid transport and insuring to the 

CEL-MAR Company.

CEL-MAR Company 

under no circumstances won't be responsible for ensuing damage from improper using the product or as a result

of random causes: the lightning discharge, the flood, the fire and the like.

CEL-MAR Company

 is not be held responsible for damages and loss including: loss of profits, loss of data, pecuniary losses ensuing

from using or the impossibility of using this product.
In specific cases

 CEL-MAR Company

 discontinue all warranties and in particular do not follow the user manual and do not accept

terms of warranty by the user.

1.2. GENERAL CONDITIONS FOR SAFE USE

The device should be installed in a safe and stable places (eg, electroinstallation cabinet), the powering cable should be arranged  so

as not to be exposed to trampling, attaching, or pulling out of the circuit. 

Do not put device on the wet surface.

Do not connect devices for nondescript powering sources,

Do not damage or crush powering wires.

Do not make connection with wet hands.

Do not adapt, open or make holes in casings of the device!

Do not immerse device in water or no other liquid.

Do not put the fire opened on device sources: candles, an oil lamps and the like.

Complete disable from the supply network is only after disconnecting the power supply circuit voltage. 

Do not carry out the assembly or disassembly of the device if it is enabled. This may result to short circuit and damage the device.

The device can not be used for applications that determine human life and health (eg. Medical). 

1.3. CE LABEL

The   CE   symbol   on   the   device   CEL-MAR   means   compatibility   with   electromagnetic   compatibility   Electromagnetic
Compatibility   Directive  

EMC   2014/30/WE.

  Declaration   of   Conformity  is   available   by   contact   with   Technical   Service

(email: [email protected]; phone: +48 41 362-12-46)..

1.4. ENVIRONMENTAL PRESERVATION

This sign on the device inform about putting expended device with other waste materials. Device should send to the

recycling. (In accordance with the act about the Electronic Appliance Expended from day 29 of July 2005)

1.5. SERVICE AND MAINTENANCE

The 

ADA-401WP module

 does not require the servicing and maintenance.

Technical support is available at 48 41 362-12-46 in 8.00-16.00, from Monday to Friday or e-mail 

[email protected]

.

1.6. PACK CONTENTS

Module is delivered with: User Manual, Line terminators 120W (2 pcs),  CD-ROM with ADAUtil software.

2. PRODUCT INFORMATION

2.1. PROPERTIES

Possibility of creating networks on the basis of RS485 bus to which are being connected 1-wire sensors,

1-WIRE to MODBUS-RTU protocol conversion,

Supporting 1-WIRE interface circuits: DS1820, DS18S20,  DS18B20, DS1822, DS2438Z, DS2401.

Temperature, humidity and atmospheric pressure measurement from the 64 digital sensors,

300m 1-WIRE bus distance - depend on sensor's number and connection method of them and cables,

Accuracy dependent on sensor's selection: for DS1820, DS18S20 – 0,50°C; for DS18B20, DS1822 – 0,0625°C,

RS-485 network baud rate up to 230,4 kbps,

1-WIRE network baud rate: Standard up to 16,3 kbps, Overdrive up to 142 kbps,

Stable power supply 10-30 VDC, power consumption up to 3W, depend on sensors number and type, 

3kV= galvanic isolation between RS-485 interface and power supply,

~3kV= opto-isolation in signal channel between RS-485 and 1-WIRE interface,

Protection against power supply reverse connection,

Integrated short circuit protection and over-voltage protection on RS-485 and 1-WIRE networks,

Screw terminal block connectors for all connections,

DIN 43880 standard - mounting in typical electro-installation unit,

Rail mounting according to DIN35 / TS35 standard,

Dimensions (W x D x H) 53mm x 62mm x 90mm.

4

ADA-401WP

Содержание ADA-401WP

Страница 1: ...User manual ADA 401WP 1 WIRE to MODBUS RTU Measuring Module 1 io_ada 401wp_v 1 24_en ADA 401WP Copyright 2001 2017 CEL MAR sp j...

Страница 2: ...1 SENSOR DISPLACEMENT 13 5 3 2 2 SENSORS EXCHANGE 13 5 3 2 3 CHANGING TEMPERATURE SENSORS ORDER BY THE USE OF AAS PROCEDURE 13 5 3 3 SENSORS REMOVING 14 5 3 4 MEASUREMENT CHANNEL PARAMETERS CONFIGURAT...

Страница 3: ...MEASURING CHANNEL 4X 3X REFERENCES 24 7 3 1 8 FUNCTION 0x03 0x04 READOUT OF CO2 CONCENTRATION MEASURING CHANNEL 4X 3X REFERENCES 24 7 3 1 9 FUNCTION 0x03 0x04 READOUT OF TEMPERATURE VALUE OF SENSOR PT...

Страница 4: ...ly of the device if it is enabled This may result to short circuit and damage the device The device can not be used for applications that determine human life and health eg Medical 1 3 CE LABEL The CE...

Страница 5: ...485 line is made on the basis of 600W protectors diodes and fuses To RS485 Bus can be connected 32 devices ADA 401WP Additional converter like ADA I1040 RS232 to RS485 or ADA I9140 USB to RS485 enabl...

Страница 6: ...erter from power circuit device that generates large impulse interference such as transmitters contactors 3 1 ASSEMBLING ADA 401WP enclosure is adapted to assembly on TS 35 DIN35 rail To install shoul...

Страница 7: ...Bd Should be used Line Termination resistor if the distance is over 1000m 9600Bd or 700m 19200Bd transmission Example connection of Rt connected to screw terminal block of RS485 422 interface are show...

Страница 8: ...500 However when building the network remember that each sensor is a shortening of 0 5 meters and 100 meters of cable causes additional capacity load data line 5nF increasing signal distortion The rea...

Страница 9: ...ecking RS485 interface Diagnosis 1 WIRE interface Diagnostic System Diagnostic Measurement channels Diagnostic 5 CONFIGURATION Configuration of ADA 401WP module should be made by the use of ADAUtil so...

Страница 10: ...disable disable manufacturer setting Errors that are signalled by short sound errors of sensors searching errors of reading temperature from sensors errors in RS485 communication errors of CRC sensor...

Страница 11: ...ingle reading sensors and Enable write TL and TH registers for DS18xxx are available from 2 005 firmware version and 1 7 1 0 version of ADAUtil 5 2 5 AUTOMATIC ASSIGN SENSORS TO MEASUREMENT CHANNELS A...

Страница 12: ...the table below Table indications of sensors Sensor Unit of measurement Sensor Unit of measurement Temperature C voltage converter 0 10V DC V U10 Relative humidity RH current converter 0 2mA DC mA A2...

Страница 13: ...iguration to module memory by pressing button Save Configuration to Module To the module memory will be saved only configuration of measuring channels which fields were marked in ping 5 3 2 3 CHANGING...

Страница 14: ...hPa to 12 70hPa for digital atmospheric pressure sensors DES 216 AP DES 300 AP etc Enable activation this field will cues actualisation of measurement and status for channel deactivation will cues blo...

Страница 15: ...in loaded file can be changed the module software Process of software update is visualized by ADAUtil in use of Progress Window and after proper update confirmed by correct message Fig 11 Choice of s...

Страница 16: ...mode value visualization of temperature measurement with the 0 5 C accuracy transmission errors on the 1 WIRE bus restore factory default and reset register Set the SW1 microswitch as in the table bel...

Страница 17: ...ter counter of Button the number of RST Reset button Register counter of Brown On specifies the number of the operating voltage drops below allowable voltage level After that can be set the switch sec...

Страница 18: ...ble below SW1 1 SW1 2 OFF OFF Yellow LED will be OFF 6 3 DIAGNOSTIC OF MODBUS PROTOCOL After finishing diagnostic of measurement channels of each ADA 401WP module in installation can be started diagno...

Страница 19: ...ters of channels 6 3 1 CONFIGURATION OF MODBUS RTU COMMUNICATION MODBUS Monitoring window has options for configuration of MODBUS RTU communication Module Address 1 247 enter the address of ADA 401WP...

Страница 20: ...sation of measurement and status for channel deactivation will cues block actualisation Assigned activation this field will cues adding a sensor to the Channel Deactivation will cues that in case of n...

Страница 21: ...condition R 16 bit register 30202 40202 1 201 Channel condition R 16 bit register 30203 40203 2 202 Channel condition R 16 bit register 30204 40204 3 203 Channel condition R 16 bit register 30204 4020...

Страница 22: ...re C 100 PT1000 sensor s temperature C PT1000 10 Relative humidity RH 10 0 10VDC transducer V U10 100 Lighting natural or unnatural Lux 10 0 20mADC transducer mA A20 100 Sunlight SLux 10 Circuit DS240...

Страница 23: ...during inquiry converting 03 CRC Lo 1 byte 04 CRC Hi 1 byte 7 3 1 2 FUNCTION 0x03 0x04 READOUT OF RELATIVE HUMIDITY FROM MEASURING CHANNEL 4X 3X REFERENCES Function 0x03 0x04 are used for readout of...

Страница 24: ...CTION 0x03 0x04 READOUT OF CO2 CONCENTRATION MEASURING CHANNEL 4X 3X REFERENCES Function 0x03 0x04 are used for readout of the value from measuring channel sensor The value from measuring channel sens...

Страница 25: ...erial number 40 0 402 Byte 4 of serial number 01 Bajt 5 of serial number 00 0 403 Byte 6 of serial number 00 Bajt 7 of serial number 23 Query of circuit serial number from channel nr 0 Byte no Designa...

Страница 26: ...status Byte Hi from ver 1 008 firmware 0 Sensor in processing 0 No 1 Yes 0 Short circuit on the 1 Wire bus 0 No 1 Yes 1 Sensor detected 0 No 1 Yes 1 DS18xx circuit present on 1 WIRE bus 0 0 No 1 Yes 2...

Страница 27: ...1 byte 83 84 02 Error code 1 byte 01 unknown function 02 unknown data address 03 unknown data value 04 unknown error while inquiry converting 03 CRC Lo 1 byte 04 CRC Hi 1 byte 8 SPECIFICATION TECHNICA...

Страница 28: ...ng Degree of casing protection IP40 Degree of terminal protection IP20 Weight 0 10 kg According to standard DIN EN50022 DIN EN43880 Location during work Free Mounting method On the rail compliant with...