HI-PE multi-zone --- Rev.
FI 050 GB 60K10 v2_1
24
Transmission parameters
Code
Meaning
Possible
values
Type
Notes
TP
Transmission
program
0-99
C/R
To choose a different program from the one pre-set in the factory, contact CEIA’s technical
department.
CH
Transmission
channel
0 - 99
C/R
0 - 49
for 50Hz power supply
50 - 99 for 60Hz power supply
Alarm signal parameters
Code
Meaning
Possible
values
Type
Notes
AD
Alarm duration
0P-5P
0C-5C
C/R
First range: proportional alarm indication
0P reset time = 0,3 s
1P reset time = 1 s
2P reset time = 2 s
3P reset time = 3 s
4P reset time = 4 s
5P reset time = 5 s
In this case the display will always provide an indication proportional to the metal mass in
transit, using the red sector on the right when there is an alarm.
Second range: constant alarm indication
0C reset time = 0,3 s
1C reset time = 1 s
2C reset time = 2 s
3C reset time = 3 s
4C reset time = 4 s
5C reset time = 5 s
In this case, during normal operation the display will indicate the selected sensitivity; when
there is an alarm, 4 asterisks light up in the red section on the right.
AV
Alarm volume
0-9
C/R
0 (disconnected alarm)
9 (maximum volume).
During local programming, the device provides an audible signal using the selected tone.
MV
Minimum alarm
volume
0-9
C/R/S 0 (disconnected alarm)
9 (maximum volume).
This determines the minimum value of the AV parameter
In the standard settings this parameter can be accessed by the superuser only
AT
Alarm tone
0-9
C/R
5 different continuous tones ranging from
0
(deep sound) to
4
(high sound)
5 different pulsed tones ranging from
5
(deep sound) to
9
(high sound).
During local programming, the device provides an audible signal using the selected tone.
Management of the Metal Detector
Code
Meaning
Possible
values
Type
Notes
BR
Serial interface
transmission
speed (Baud
rate)
110, 150,
300, 600,
1200,
2400,
4800,
9600
C/R
The change of serial transmission speed becomes operative when the programming
session is terminated using the
PE
command.
NP
New password
XXXXXX
X = any
alpha-
numeric
character
C
This selects a new password to allow access the programming phase. The password must
be made up of
6 letters
(either upper or lower case) and/or numbers. The password is
displayed to facilitate entry.
Once the new password has been entered, it will be
effective immediately.
ATTENTION!
For security reasons, once a password has been entered it will no longer be
possible to read it. Should the access code be forgotten, assistance will be required from
C.E.I.A. personnel.
AC
Number of
alarms detected
0-999999 R
Model HI-PE
with photocells
It displays:
•
the number of passages through the antenna (
P
, 6 figures),
•
the number of passages with alarm (
A
, 6 figures)
•
the net alarms/passages ratio as a percentage (
RR
-- ring rate, 2 figures and one