TROUBLESHOOTING

PROBLEM

CAUSE

SOLUTION

The mixer cannot be 

powered on

Power cord is disconnected or

main power switch is off

Plug the power cord in again and

check the main power switch

The fuse is blown

Inspect the fuse and change if

needed

Abnormal sounds 

or vibrations during 

operation

Motor screws at the base of the

unit haven’t been removed

Remove the motor fixing screws 

as instructed in page 5

Operation of the mixer without

mixing cup or with empty cup

Stop operation

Lid of mixing cup is loose

Stop operation and completely

close the lid

Defective gears

Return to the dealer for repair

Display panel does

not light up

Defective CPU Board

Return to the dealer for repair

Unit operation has

delayed start

Defective CPU Board

Return to the dealer for repair

Mixing cup not closing 

or opening well

Mixing cup deformed

Replace mixing cup

Blocked air hole in the lid

Free the air hole

Mixing results are poor

Incorrect mixing ratio of alginate 

and water, or incorrect mixing 

time

Refer to instructions from

alginate supplier

Incorrect operating steps

Refer to Panel Operations

Section C on page 6

Poor or expired alginate

powder

Use a new batch of alginate

powder

Deteriorated rotation speed

Return to the dealer for repair

Display panel shows 

‘E1’

The lid of the mixer isn’t

completely closed

Open and fully close the lid again

Defective microswitch

Return to the dealer for repair

Display panel shows 

‘E2’

The lid buckle of the mixer does 

not work

Open and fully close the lid again

Defective microswitch

Return to the dealer for repair

Display panel shows 

‘E3’

Unstable operating platform

Ensure operating platform is

stable

Defective microswitch

Return to the dealer for repair

When requested to send the unit back for repair, please send the full unit and power cord for 

further inspection.

ENGLISH

7

Содержание Alginate Mixer 3

Страница 1: ...CAVEXYOUR IMPRESSION IS OUR CONCERN www cavex nl User manual Cavex Alginate Mixer 3 Installation Operation Troubleshooting ...

Страница 2: ...the Alginate Mixer 3 in the manner specified below using the Cavex Alginate Mixer 3 for materials other than alginate impression material or gypsum or in a manner other than that described below using parts other than those supplied for the intended purpose by Cavex or repairs or modifications other than those made by experts appointed by Cavex for that purpose SAFETY PRECAUTIONS Please follow the...

Страница 3: ...er States Explanation of Graphical Symbols IEC 60878 and ISO 3864 B 3 6 Warning Dangerous voltage ISO 7000 0434 Caution consult accompanying documents ISO 7000 1641 Follow operating instructions or consult instructions for use IEC 60417 5009 Stand by ISO 7010 W001 General warning sign MAINTENANCE Alginate Mixer If there is any dust or stains on the surface of the control panel or unit body please ...

Страница 4: ...on of the manufacturer 1 Power button Green indicator Ready for operation Orange indicator Power on 2 Display 3 Start stop button 4 Main power switch 5 Fuse backup fuse 6 Power cord port 7 Power cord 8 Cavex Alginate Mixer 3 9 Spatula 10 Powder scoop 11 Stirrer stick 12 Water dosing bottle 13 Accessory holder 14 Mixing cup m m 4 2 1 3 14 12 13 7 11 9 6 4 10 8 5 ...

Страница 5: ...tor fixing screws 1 Remove the two motor fixing screws located at the base of the unit 2 Store the two screws and distance washers in the openings at the back of the unit for future use IMPORTANT If the unit needs to be moved or sent for repairs secure the motor with the two motor screws prior to transporting it to avoid damages IMPORTANT Keep the unit upright when removing installing the screws t...

Страница 6: ...p has been securely closed 2 Place the prepared mixing cup firmly into the mixing cup holder 3 Close the lid of the Cavex Alginate Mixer 3 and check that the appropriate mixing time has been set 4 Press the Start stop button to start the mixing cycle 5 Wait for the stop beep signal before opening the lid 6 Remove the alginate mixture from the mixing cup with a spatula and apply CHARGER SIGNALS CAU...

Страница 7: ...well Mixing cup deformed Replace mixing cup Blocked air hole in the lid Free the air hole Mixing results are poor Incorrect mixing ratio of alginate and water or incorrect mixing time Refer to instructions from alginate supplier Incorrect operating steps Refer to Panel Operations Section C on page 6 Poor or expired alginate powder Use a new batch of alginate powder Deteriorated rotation speed Retu...

Страница 8: ...5 100 g cup Dimensions 273 L x 219 W x 309 H mm Net weight 16 8 kg Power input AC 220 240V Electric shock protection level Class I Fuse Rating 250V 6 3A Cavex Holland BV Fustweg 5 2031 CJ Haarlem The Netherlands T 31 23 530 77 00 dental cavex nl www cavex nl ZYAT545 01GB 00 June 2019 CAVEXYOUR IMPRESSION IS OUR CONCERN www cavex nl ...

Отзывы: