Extraordinary maintenance must be entrusted to a trained
technician in possession of original spare parts.
• Pay special attention to danger signals. Use protective
goggles, gloves and disposable overalls for personal pro-
tection.
• Check that routine maintenance has been duly carried out
and make sure that Magnolia products are used.
• Before any intervention carry out a series of washes with
Fast&Steril 3
(D)
disinfectant, then wait 15 minutes for a
complete disinfectant action.
• Check maximum detected temperatures and all alarms.
Intervene accordingly.
• Where routine maintenance is not carried out properly
or unsuitable products are used, train the staff and inform
the person in charge. Warranty ceases in case appliances
are treated with products which are different from those re-
commended.
• Check the aspirator noise level (see page 26).
• Remove dust from the control panel’s fan and heat sink
using a blast of dry air not exceeding 2 bar pressure. By
means of a 6 bar blast of air clean also the small holes on
the frontal cover of the aspirating unit
(15)
.
• Check the conditions of plastic hoses, in particular the
hoses under pressure connecting the centrifugal separator
(11)
and the Hydrocyclone ISO. We suggest to replace the-
se hoses every 18-24 months.
• Check the working conditions of the centrifugal separator
(11)
and re-circulation valve
(14)
.
• Every time a component containing a rubber part (“O”
ring, gasket or seal) is disassembled, replace the rubber
part.
• Replace the motors bearings.
(*) Suggested maintenance intervention frequency is approximate. In
case of works causing a lot of dust or particularly intensive, increase
the frequency of interventions according to the needs.
Recommended every 12 months *
Recommended every 18-24 months
Recommended every 10,000/15,000
working hours
B
D
32
• Extraordinary maintenance
ALLARME:
TEMPERATURA ELEVATA
15
11
14