The Pro DX Sound Level Meter can be set so that once a certain sound level 
has been reached the DAT machine will start recording. 
 
Once the sound level falls below this level the DAT machine will then stop re-
cording. 
 
To set this level and to enable the DAT option press the 

menu

 key on your Pro 

DX Meter until the Main Menu is displayed. 
 
Select option 2 ‘General Setup’ by pressing key 

2. 

 

Select option 4 ‘DAT Recorder’ by pressing key 

4

 

Select option 1 ‘DAT activated’ by pressing key 

1

 
Press the Light Blue Arrow key directly underneath TOGGLE  displayed on the 
screen.  This will toggle the DAT to be either ON or OFF.  (Note YES = ON).  
Press the 

OK

 button to confirm the selection. 

 
Select option 2 ‘DAT threshold’ by pressing key 

2

 
Type in a number between 30 and 140 (Note that this figure needs to be less 
than or equal to the top of the range you have selected.  As an      example if 
you have selected the 15 to 90dB range then do not set this value to greater 
than 90dB.  Press the 

OK 

button once your threshold is entered. 

 
Repeat press the Light Blue Arrow key directly underneath BACK that is         
displayed on the screen to return to sound level meter operation. 
 
The Pro DX Meter is now setup to automatically start the DAT recording at 
your selected threshold level when you start the Pro DX to record.  To per-
form this task press the Start / Stop key which is located to the left of the OK 
button.  To stop the Pro DX Meter recording simply press the Start / Stop 
key again.  Note that REC is displayed on the screen whilst the Pro DX Meter 
is recording. 

 

The DAT records for a minimum of 5 seconds even if the sound level has 
dropped below the set threshold level.  This delay is reset each time the sound 
level reaches the threshold level.  The purpose is to alleviate a   recording that 
continually starts and stops if the noise source is          fluctuating around the 
set threshold level. 

12 

P

RO

 DX S

ETUP

 F

OR

 DAT R

ECORDING

 

DAT 

 

To ensure validity of the recorded data it is necessary to record a reference 
tone on the tape each time your KA018 kit is used. 
 
Follow the procedure below to perform this task:  
 
Firstly ensure that your microphone preamplifier is plugged into your 
microphone extension cable, then plug the microphone extension cable into 
the socket labelled ‘microphone’ on the KA018 case. 
 

TAKE EXTREME CARE WHILST HANDLING YOUR MICROPHONE, IT IS 

AN EXTREMELY SENSITIVE DEVICE AND VERY COSTLY 

 
Ensure that your Pro DX Meter and the DAT are ON.  (If the DAT machine is 
currently OFF it can be powered by pressing the 

STOP

 key) 

 
If you have not yet set the DAT clock, do so now as indicated on page 8 of 
this manual. 
 
Ensure the range 45 – 120 is selected on your Pro DX Meter, this is 
achieved by pressing the Light Blue Arrow key directly underneath RANGE 
on the screen until the desired range is selected.  (If RANGE  is not visible 
then repeat press 

menu

 until it is) 

 
Confirm that the 3.5mm jack plug is connected to the DAT Line Input socket 
and then press the  

REC / ID WRITE 

key on the DAT machine. 

 
Carefully and gently push a Castle GA607 calibrator over the microphone.  A 
small amount of resistance should be felt  as the o-ring seal on the 
calibrator forms a seal around the microphone.  
 
Ensure the calibrator is set to 94dB. (Note that the calibrator automatically 
turns OFF after approximately 1 minute) 
 
Set the 

REC MODE 

to manual on the DAT machine and then adjust the dial 

on the DAT machine labelled 

REC LEVEL 

until the both the left and right 

channels  read  –24dB  (minus  24dB)  on  the  bar  graph.    i.e.  until  the  bar 
graph display is located directly above the number 24. 
 
Switch the calibrator to 94dB again to reset its timer to give another 
minutes operation.  Press 

PAUSE

 on the DAT machine twice to start 

recording the reference tone for future playback and analysis purposes. 

 

DAT R

EFERENCE

 T

ONE

 

DAT 

Содержание KA018

Страница 1: ...HB 1018 006 A5 Issue C NOTES...

Страница 2: ...d like to become a competent person in the eyes of the law then you may like to know more about our Competent persons training courses which cover a wide variety of noise assessment tasks It is my int...

Страница 3: ...tteries in the KA018 if it is not in use Do not let any conductive objects such as wire or metal scraps get into the unit Care should be taken when closing the case lid to ensure that no cables are tr...

Страница 4: ...al Options 18 Trouble Shooting 18 Warranty and After Sales Service 19 Accessories 20 Technical Specifications 21 Notes 22 TECHNICAL SPECIFICATIONS Operating Voltages And Current Consumption GA131 Soun...

Страница 5: ...rproof Case is part of Castle s Pro DX meter accesso ries and has been designed to allow outdoor and long term logging capabilities with optional sound recording to a Digital Audio Tape DAT recorder T...

Страница 6: ...to the DAT Example Title DAT 2 Warranty and After Sales Service Castle Group Ltd design and manufacture precision instruments which if treated with reasonable care and attention should provide many y...

Страница 7: ...ng Ensure the printer is ONLINE indicated by a solid GREEN LED on the top face of the printer The solid RED LED indicates the printer is OFFLINE Press the ONLINE OFFLINE key to toggle between ONLINE o...

Страница 8: ...als as required Ensure that the 12V batteries are connected or alternatively if you wish to power the KA018 kit from the mains plug the power lead into the socket labelled mains or ext battery Plug yo...

Страница 9: ...e the Battery Disconnect Switch is ON indicating Blue and insert the charging plug into the control box socket labelled Charge Charging both batteries completely if depleted may take approximately 14...

Страница 10: ...l still be possilbe The supplied GA505 printer will operate using the power supplied from your KA018 kit as with your Pro DX Sound Level Meter The printer is only powered when print information is sen...

Страница 11: ...SENSITIVE DEVICE AND VERY COSTLY Press the Power key on you Pro DX meter to turn on your instrument Once powered up you are presented with an option to calibrate your instrument by selecting YES NO P...

Страница 12: ...conds 2 Press the or buttons to set the current year Press the CLOCK SET button when the correct year is selected 3 Repeat step 2 to set the current Month Date Day Hour Minute and Seconds The display...

Страница 13: ...d stops if the noise source is fluctuating around the set threshold level 12 PRO DX SETUP FOR DAT RECORDING DAT To ensure validity of the recorded data it is necessary to record a reference tone on th...

Страница 14: ...depth information on range selection 10 DAT REFERENCE TONE CONTINUED DAT PRO DX RANGE SETUP SOUND METER A microphone hears all frequencies equally whereas a human ear hears some frequencies better th...

Отзывы: