The KA018 kit may be powered from either the supplied Castle 12V DC Power 
Supply or from an optional external Castle Battery Pack. 
 
To use any external power source simply plug it into the ‘Mains / Ext Batt’ 
Connector on the KA018 case and remove when finished.  This procedure will 
not interrupt your instruments operation if your batteries are charged and 
connected correctly. 
 
The position of the battery isolation switch will not hinder operation with either 
the external  12V DC Power Supply or the optional External Batteries, however 
the isolation switch will need to be ON for un-interruptible operation between 
Internal Batteries and External Supply. 
 
For safety reasons only use the supplied Castle DC Power Supply and optional 
Castle Battery Pack. 
 
 
 

The power switch on the printer can be left on at all times and the printer’s 
internal battery can be removed when used in a KA018. 
 
With the GA505 printer plugged into the KA018, power is automatically 
provided to the printer when data is sent from the instrument. This means that if 
the batteries are not installed in the printer or if they are flat, printing will still be 
possilbe. 
 
The supplied GA505 printer will operate using the power supplied from your 
KA018 kit as with your Pro DX Sound Level Meter.  The printer is only powered 
when print information is sent to the printer.  When the print job is finished the 
printer will power down again.  
 
To  ensure  printing  runs  smoothly  ensure  that  the  power  supply  and  9  pin  D 
connector are plugged into the printer.  
 
Printing is achieved via your Pro DX Sound Level Meter,  refer to your operating 
manual if required. 

E

XTERNAL

 B

ATTERY

 P

ACK

 &  P

OWER

 S

UPPLY

 

A

LL

 

P

RINTING

 

P

RINTER

 

 

The MW401 is an optional longer term weather proof windshield assembly 
with a de-humidifier unit that can be used for periods of time greater than 7 
days.  The unit is assembled using the additional microphone pre-amplifier 
that has been supplied with your MW401 kit.  Use the diagrams below to aid 
assembly of the unit.  Please remove your microphone capsule prior to 
assembly. 
 
 
 
 
 

A

TTACHING

 T

HE

 M

ICROPHONE

 

TO

 

THE

 MW401 

W

INDSHIELD

 

Standard 

Pre-Amplifier 

Additional Long Term 

Pre-Amplifier 

De-Humidifier Unit 

Step 1 

Attach the long term pre-amplifier unit to the microphone extension cable. 

Step 2 

Insert the pre-amplifier through the MW401 stem and attach the de-humidifer 
unit to the top of the pre-amplifier.  Do not attach the microphone capsule at this 
stage. 

15 

Содержание KA018

Страница 1: ...HB 1018 006 A5 Issue C NOTES...

Страница 2: ...d like to become a competent person in the eyes of the law then you may like to know more about our Competent persons training courses which cover a wide variety of noise assessment tasks It is my int...

Страница 3: ...tteries in the KA018 if it is not in use Do not let any conductive objects such as wire or metal scraps get into the unit Care should be taken when closing the case lid to ensure that no cables are tr...

Страница 4: ...al Options 18 Trouble Shooting 18 Warranty and After Sales Service 19 Accessories 20 Technical Specifications 21 Notes 22 TECHNICAL SPECIFICATIONS Operating Voltages And Current Consumption GA131 Soun...

Страница 5: ...rproof Case is part of Castle s Pro DX meter accesso ries and has been designed to allow outdoor and long term logging capabilities with optional sound recording to a Digital Audio Tape DAT recorder T...

Страница 6: ...to the DAT Example Title DAT 2 Warranty and After Sales Service Castle Group Ltd design and manufacture precision instruments which if treated with reasonable care and attention should provide many y...

Страница 7: ...ng Ensure the printer is ONLINE indicated by a solid GREEN LED on the top face of the printer The solid RED LED indicates the printer is OFFLINE Press the ONLINE OFFLINE key to toggle between ONLINE o...

Страница 8: ...als as required Ensure that the 12V batteries are connected or alternatively if you wish to power the KA018 kit from the mains plug the power lead into the socket labelled mains or ext battery Plug yo...

Страница 9: ...e the Battery Disconnect Switch is ON indicating Blue and insert the charging plug into the control box socket labelled Charge Charging both batteries completely if depleted may take approximately 14...

Страница 10: ...l still be possilbe The supplied GA505 printer will operate using the power supplied from your KA018 kit as with your Pro DX Sound Level Meter The printer is only powered when print information is sen...

Страница 11: ...SENSITIVE DEVICE AND VERY COSTLY Press the Power key on you Pro DX meter to turn on your instrument Once powered up you are presented with an option to calibrate your instrument by selecting YES NO P...

Страница 12: ...conds 2 Press the or buttons to set the current year Press the CLOCK SET button when the correct year is selected 3 Repeat step 2 to set the current Month Date Day Hour Minute and Seconds The display...

Страница 13: ...d stops if the noise source is fluctuating around the set threshold level 12 PRO DX SETUP FOR DAT RECORDING DAT To ensure validity of the recorded data it is necessary to record a reference tone on th...

Страница 14: ...depth information on range selection 10 DAT REFERENCE TONE CONTINUED DAT PRO DX RANGE SETUP SOUND METER A microphone hears all frequencies equally whereas a human ear hears some frequencies better th...

Отзывы: