background image

CASELLA USA 

 

 

 

CEL-240 DIGITAL SOUND LEVEL METER 
FREQUENTLY ASKED QUESTIONS

 

 

Casella USA 

Page 6 of 6 

(800) 366-2966 

31 Jan 2008 

[email protected]

 

FAQ’s:  ACCURACY & RESULTS 

 

What accuracy is the instrument designed to fulfil? 

The 

CEL-240

  instrument  complies  with  the  international  standards  in  the type  2  classifications that 

classify  the  sound  level  meter  as  general-purpose  instruments.  This  specifies  an  overall  expected 
accuracy of +/- 2 dB under normal measurement conditions. Use of the instruments as dosimeters by 
visually averaging fluctuating noise levels may produce wider tolerances than this depending on how 
well the “on time” of a notional average level is measured. 
 

Are any results saved when using the SLM mode? 

Results are not saved in the instrument when any of the available display modes are used. The 

CEL-

240

 is intended to allow the instruments to be used as a simple sound level meter for quick hand held 

surveys or as a simple front end to a data logging system in conjunction with a personal computer 
running the optional dB24 software.  

 

FAQ’s:  APPLICATIONS 

 
What markets is the instrument designed for? 

There are three main markets for the 

CEL-240

a simple meter for monitoring noise exposure in the workplace 

a general purpose sound level meter for many environmental nuisance sources 

a short term noise meter for quick measurements on specific noise sources 

 
How are these markets catered for by the CEL-240? 

For  many  basic  noise  measurements  the  A  weighted  Slow  response  sound  level  will  be  all  that  is 
required.  For more  specific  noise measurements  of  sounds  with  a  transient  nature the  C  weighting 
and the maximum hold feature will be of help. The meter can also be used as a simple teaching aid 
in many noise or general sound studies especially with the optional digital output to a computer. The 
tripod mount allows the meter to be fixed at a chosen location for longer term measurements that are 
not likely to be adversely affected by the weather conditions. 
 
 
 
 
 
For more details on the 

CEL-240

 series, or any of the other Casella USA products, please contact 

us: 
 

Casella USA 
17 Old Nashua Road, #15 
Amherst, NH 03031 
(800) 366-2966 

[email protected]

 

www.CasellaUSA.com

 

 
 

Содержание CEL 240 Series

Страница 1: ...trument be powered from an external source Can the microphone of a CEL 240 be removed Can the CEL 240 be used for outdoor measurements How quickly does the digital display update during measurements Is the time history trace stored in the CEL 240 What outputs are available from the CEL 240 What are the characteristics of the ac output What are the characteristics of the dc signal What are the char...

Страница 2: ...vice versa Both models feature ANSI type 2 accuracy for overall measurements The new CEL 350 dBadge dosimeter adds a small badge style dosimeter with cable less microphone to the range for added convenience CEL 400 Series The mid range sound level meters are the CEL 430 450 and 490 providing recording of level against time or level against frequency The 450 and 490 instruments are fitted with para...

Страница 3: ...s are recommended for longest running time Can rechargeable batteries be used in a CEL 240 A set of 3 x AA NiMH rechargeable batteries can be used instead of alkaline batteries and battery life will be around the same time at normal room temperature The rechargeable batteries will operate successfully down to lower temperatures than an alkaline battery set NiMH batteries must be recharged outside ...

Страница 4: ... The C setting is recommended for making audio recordings since it has a better wider and flatter frequency response than the A weighting The ac signal is available on the 2 5 mm jack socket at a level of up to 0 85 V rms for full scale deflection This means that for a sound level of 100 dB measured on the Lo range of the meter the output would be 0 85 V rms A signal at half the range FSD would ou...

Страница 5: ...er to see what is happening Use the Fast response to more accurately follow the changing noise levels as they rise and fall Use the Impulse response only when the regulations require it What exchange rates or Q values are available in the CEL 240 There is no specific exchange rate available in the meters since this function is only associated with instruments that can measure the time average nois...

Страница 6: ... with a personal computer running the optional dB24 software FAQ s APPLICATIONS What markets is the instrument designed for There are three main markets for the CEL 240 a simple meter for monitoring noise exposure in the workplace a general purpose sound level meter for many environmental nuisance sources a short term noise meter for quick measurements on specific noise sources How are these marke...

Отзывы: