MAXTRAP 50 & 105 Supplement Instruction Manual Rev 11.29.18

www.cascadesciences.com | [email protected] | 503-847-9047

Installation

Pre-Installation

Check that the ambient condition, ventilation, and spacing requirements for the Thermo refrigerated 

unit are met – See Thermo manual , page 20  (Allow at least 12” on the right side of the unit for venti-

lation and air to draw thru the  slotted right side of the Thermo unit. 

Electrical: 120V

The 

Labconco refrigerated base unit 

 requires a standard 110 – 120 volt NEMA 5-15 power supply 

outlet to plug into.

The

 220V version

 requires a NEMA 6-15R 220-240V power supply outlet to plug into.

Please refer to the oven data plate for individual electrical specifications. 

Technical data specified applies to units with standard equipment at an ambient temperature of 25°C 

(77°F) and a voltage fluctuation of ±10%.

Содержание MAXTRAP 105

Страница 1: ...s MAXTRAP 50 105 Supplement Instruction Manual Rev 11 29 18 www cascadesciences com info cascadesciences com 503 847 9047 Supplemental Installation Operation Manual MAXTRAP 50 105 Refrigerated Cold Tr...

Страница 2: ...protection impairment by disabling or interfering with unit safety features This can result in damage to the unit and injury death or negative effects Follow all national laws regulations and local o...

Страница 3: ...llowing information ready when calling or emailing Technical Support The model number and the serial number 503 847 9047 info cascadesciences com Cascade Sciences 6725 NE Evergreen Pkwy 106 Hillsboro...

Страница 4: ...e period otherwise Cascade Sciences will not be held responsible Take photos of any shipping damage Refuse delivery or make note on the bill of lading 1 If anything is missing please call us at 503 84...

Страница 5: ...separately from the MAXTRAP It is strongly recommend that the vacuum pumps GAS BALLAST be in the open position to best tolerate ethanol and other gas loads In some cases the vacuum pump can be powered...

Страница 6: ...2 on the right side of the unit for venti lation and air to draw thru the slotted right side of the Thermo unit Electrical 120V The Labconco refrigerated base unit requires a standard 110 120 volt NEM...

Страница 7: ...OVEN Make sure your have 800 ml of a heat transfer fluid such as ethanol glycol or a cryocool inside the thermo reservoir This fluid is sold separately 1 Remove the stainless COLDWELL Pour the 800ml o...

Страница 8: ...l as important notes in the user manual Symbol Definition Consult the user manual Consulter le manuel d utilisation AC Power Rep re le courant alternatif AC Power Rep re le courant alternatif Begin he...

Страница 9: ...not have to be under vacuum to operate Allow at least 30 minutes for the MAXTRAP fluid to reach operating temperature From The Vacuum Oven Vacuum valve is closed Vent Valve is closed Product is loade...

Страница 10: ...ing Precautions 1 Do not move during operation 2 SAFETY TIP At very cold temperatures solvents and even oxygen can re condense back into liquid This is potentially explosive Especially if the cold tra...

Страница 11: ...s that contain solvents capable of harming paint coatings or stainless steel surfaces Do not use chlorine based bleaches or abrasives these will damage the COLDWELL liner Vacuum Pump Maintenance Vacuu...