9
2.1. How to Go to Normal Mode (C – 0)
To go to calibration mode, turn on the power while pressing the calibration switch. The
display shows “CAL” three times, and then “C-0”.
To go to Normal mode, press the [ZERO] key.
2.2 How to Confirm Span Calibrated A/D Value(C – 1)
(1) To go to calibration mode, turn on the power while pressing the calibration switch. The
display shows “CAL” three times, and then “C-0”.
(2) Press the [TARE] key until the display shows “C – 1” .
(3) Press the [ZERO] key, the display shows “0”.
(4) Place full capacity on the platter. The display shows “90,000”. If the number is not “90,000”,
press the lb/kg(*) or TARE key to have number up or down.
(5) Press the [ZERO] key, the display shows “C – 1”. You can change the mode by pressing
TARE or lb/kg(*) key.