Simple calibration without weight 

1. Turn on the scale, and key in “000419” during counting down (self-check) to zero to enter into Simple Calibration mode. 

The displays will indicate as below.   
Press 

MOVE

 key to choose the taget platform(local or remote).   

 

WEIGHT  

UNIT 

WEIGHT 

 

COUNT 

 

CAL 

locAL 

 

------

 

▼ ▼  

▼   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NET    g 

Kg 

Ib 

 

 

 

HI

OK

LO

   

 

NEGATIVE Local Remote
COUNT

 

Then press 

ENTER

 key for determination and enter into calibration mode.   

 

    2. The default detting is 1/3 capacity(Take 6kg scale for instance) 

 

WEIGHT  

UNIT 

WEIGHT 

 

COUNT 

 

Unit 

 

------

 

▼ ▼  

▼   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NET    g 

Kg 

Ib 

 

 

 

HI

OK

LO

   

 

NEGATIVE Local Remote
COUNT

 

Note: The weighing unit for Local platform is g, while the unit for Remote platform is kg. 

3. Put a weight on the pan same as what exactly shown in the   

UNIT WEIGHT window, then press the

 ENTER

 key to confirm the operation. 

The displayed reading in the UNIT WEIGHT window starts blinking.The scale will stop blinking and return to normal counting mode. 

Calibration is now completed. 

Note: 

 

Press

 CLEAR 

key to escape from calibration mode at any time.

 

 

Change calibration value 

After entering the third step, press 

MOVE

 key. Use numeric keys to input a calibration value r( 0.80000~ 1.20000). Press 

ENTER

 key to confirm, then 

the calibration is finished.

 

When the scale was calibrated in Country A, and it is used in Country B, the error coefficient (R) should be: 

             

       R=gravity value in Country A / gravity value in Country B 

g

g

Содержание EC-II

Страница 1: ...ay data doesn t change or the internal resolution value is not in the natural range c The internal resolution data drift badly d The initial resolution is drifting 3 Trouble of power supply The phenom...

Страница 2: ...ce it Check the output voltage of adapter is about 17V The AC jack for adapter is damaged replace it The adapter is damaged Please replace it Check if there is power output from the audion T5 Check if...

Страница 3: ...age value between the pins of oscillator CR1 is 0 20v Change the oscillator Replace the CPU The main board is damaged Please replace it Displays well when power on The display is faulty Replace the 16...

Страница 4: ...en pressing the keys To short circuit the keypad The problem can be solved The membrane keypad is damaged Change the 16 cores wire The problem can be solved The 16 cores wire is damaged Change the CPU...

Страница 5: ...recalibrate The problem can be solved The load cell is damaged Check if the working environment is stable The voltage value is normal The battery voltage is less than 5 4V Recharge or replace the bat...

Страница 6: ...e problem can be solved The 9 cors wire is damaged Change the display board The problem can be solved The problem is in display board Check if all the pins were welded bad Reweld them The problem is i...

Страница 7: ...ain board and load cell is changed the scale must be recalibrated RS232 don t transmit any data Check the voltage of the 7 pins of the RS232 socket is around 10V The socket is damaged replace it Check...

Страница 8: ...calibrated before WEIGHT UNIT WEIGHT COUNT 0 0000 00 200358 NET g Kg Ib HI OK LO NEGATIVE Local Remote COUNT 1 2 1If you want to change the calibration setting values please ENTER key 1 2 2If there i...

Страница 9: ...n Count display window shows the total internal resolution which equals to original one and the increasing value Press SET key again to back to calibration mode 3 Change and choose the unit resolution...

Страница 10: ...COUNT 3 2 2 Press the MOVE key to change the value kg Ib 3 2 3 Press the ENTER key to confirm and enter into next step Choose Resolution 3 3 Choose Resolution value 3 3 1 The display shows WEIGHT UNI...

Страница 11: ...alue 1 2 5 3 4 3 Press the ENTER key to confirm and enter into next step Choose capacity 3 5 Choose capacity 3 5 1 The display shows WEIGHT UNIT WEIGHT COUNT HHHHH H CAP 200358 NET g Kg Ib HI OK LO NE...

Страница 12: ...he ENTER key to confirm and enter into next step Enter the first calibration value 3 7 Enter the first calibration value 3 7 1 The display shows WEIGHT UNIT WEIGHT COUNT HHHHH H CAP1 200358 NET g Kg I...

Страница 13: ...display window will show End Then press ENTER key to back to 1 2 of step1 without saving the second calibration value 3 9 Enter the third calibration value 3 9 1 The display shows WEIGHT UNIT WEIGHT...

Страница 14: ...value Put on the weight refer to the display Press SMPL key for determination If SET key is pressed in step 3 9 press SMPL key calibration is finished and back to normal mode z Several seconds later t...

Страница 15: ...r Local platform is g while the unit for Remote platform is kg 3 Put a weight on the pan same as what exactly shown in the UNIT WEIGHT window then press the ENTER key to confirm the operation The disp...

Отзывы: