4. Calibration
Mode:
z
Press
“
SMPL
” key to do the calibration. The unit weight window shows”00”, and the value”HHHHHH” in weight window blinks.
z
Several seconds later,the unit weight window shows “ 01 “,and the weight window shows the weight value (the first calibration value). Put on the weight refer
to the display. Press “
SMPL
” key to confirm.
If “
SET
” key is pressed in step
3.8
, press “
SMPL
” key, calibration is finished and back to normal mode.
z
Several seconds later,the unit weight window shows “02”, and the weight window shows the weight value (the second calibration value). Put on the weight
refer to the display. Press”
SMPL
” key for determination.
If “
SET
” key is pressed in step
3.9
, press “
SMPL
” key, calibration is finished and back to normal mode.
z
Several seconds later,the unit weight window shows “ 03” ,and theweight window shows the weight value (the third calibration value).Put on the weight refer
to the display. Press “
SMPL
” key to confirm.
z
Several seconds later,the scale return to counting modle.Then the caliberation procedure is finished.
z
Switch off, and take off all the weight, then swithc on, the scale can be used.
Note:
If the scale hasn’t be calibrated yet, value in weight display window is random data except unit and accuracy. On the contrary, value in weight window
should be calibration data or last keyed in ones.