6200 series Operations Manual
Rev D 2/14/2012
Page 18 of 60
Installing the Heatless Dryer Package (DRYR301)
The Heatless Dryer Package extends the operational limits to a minimum humidity
control point of 2% RH and improves the dehumidification rate by purging the chamber
with dry air. Air flows through the Dryer Package only when dehumidification is
required to maintain the humidity set point.
It uses compressed air at 90 to 100 psig to operate. Maximum airflow required is 480
SCFH (8 CFM). Of this 300 SCFH (5 CFM) is injected into the chamber as dry air and
180 SCFH (3 CFM) is exhausted into the room after being used to regenerate the
desiccant. Particles of dirt or rust, as well as large amounts of condensed moisture or
oil that may be in the compressed air line feeding the dryer must be removed from the
air stream before it enters the dryer.
Air Supply
Inlet
Tower Clamps
(optional)
Power Cord
Air Tube
Coalescent Filter
With Drain
Pressure
Regulator Dial
Pressure
Regulator Knob
Flow Meter
Adjustment
Mounting
Brackets
Flow Meter
Outlet
Dryer Panel