22
Restarting the system
As with any other computer-controlled device, it is possible for the
software in your CARO system to “crash”. This may be caused by
external influences, for example strong electromagnetic interference, or
previously undetected software errors.
If you think that your CARO system is responding incorrectly or it is
not responding at all, then we recommend restarting the microprocessor.
There are different ways to do this:
-
If the fuse in the wiring harness is easily accessible, remove the
fuse and then re-insert it into its holder after a few seconds.
-
If the fuse is not accessible, disconnect the Western connector on
the right-hand side of the control panel. To do this, press the
plastic clips from behind against the body of the connector and
then pull the connector out from the unit. Next, press the two
buttons
and
on the control panel, keep them pressed and
plug the Western connector back in again. Now you can let go of
the two buttons.
In both of the cases described above your CARO system will perform
an internal restart. If the system is unfolded, it is retracted straight away.
After the restart the system is in rest mode and can be started in the
usual way using the
key.