6
Chapter 3. Introduction of NGA 6000
3-1 Specification
NGA 6000
Measuring item
CO, HC, CO2, O2, Lambda(air surplus rate),
AFR, NOX (optional)
Measuring
method
CO, HC, CO2 : NDIR Method
O2, NOx : Electrochemical Cell
Measuring range
CO
0.00 ~ 9.99%
HC
0 ~ 9999 ppm
Resolution
0.01%
1 ppm
Display
4 digit 7segment LED
4 or 5 digit 7segment LED
Measuring range
CO
2
0.0 ~ 20.0%
O
2
0.00 ~ 25.00 %
Resolution
0.1%
0.01 %
Display
4 digit 7segment LED
4 digit 7segment LED
Measuring range
Lambda
0 ~ 2.000
AFR
0.0 ~ 99.0
Resolution
0.001
0.1
Display
4 digit 7segment LED
4 digit 7segment LED
Measuring range
NOx
(Option)
0-5000 ppm
Resolution
1 ppm
Display
4 digit 7segment LED
Repeatability
Less than ±2% FS
Response time
Within 10 seconds (more than 90%)
Warming up time
About 2 ~ 8 minutes
Sample collecting
quantity
4 ~ 6 L/min
Power
220V AC or 110V AC ±10% 50 / 60Hz
Power consumption
About 50 W
Operation
temperature
0 ~ 40
℃
℃
Dimensions
420 (W) × 298 (D) × 180 (H) mm
Weight
About 6.9 kg
Basic accessories
Probe, Probe hose, Spare fuse, Leak test cap,
Spare filter, Operation manual, Power cord,
RS232, Communication cable, Printer, Printer paper